DQR Information
D110927.3 [nsa/gvrp/C1]
DQR Submitter
Maria Cadeddu
Subject
NSA/GVRP/C1 - High noise in the data
Description
All channels of the radiometer became suddenly very noisy. The receiver noise figure was computed and was found anomalous. Data are unusable during the indicated times. On 10/13 the instrument was removed and will be sent to the vendor for repairs.
Suggestions
Do not use.
Affected Time Spans
Start Date/Time | End Date/Time | Data Quality Metric |
---|---|---|
2011-08-01 00:00:00 | 2011-10-13 22:00:00 | Incorrect |
Measurements
nsagvrpC1.b1(1)
- brightnessTemperature
DQR Attachments
-
No attachments found!