DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D161007.4 | SGP/AIP/C1 - Wet NEPH humidifier removed | sgpaipfitrh1ogrenC1.c1 | Missing |
D120703.1 | SGP/AOS/C1 - Missing data | sgpaoscpcC1.a1 | Missing |
D160615.4 | SGP/AIP/C1 - NEPH T/RH values from thermal gradients under review | sgpaipfitrh1ogrenC1.c1 | Suspect |
D160715.4 | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing | sgpaosacsmC1.a1 | Note |
D161128.1 | SGP/AIP/C1 - Reprocess: NaNs in Data | sgpaipfitrh1ogrenC1.c1 | Note |
D160921.2 | SGP/TDMA/C1 - Missing Data | sgptdmahygC1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-07-13 | 17:43:00 | 2016-07-28 | 15:00:00 | Missing |
Subject: | SGP/AIP/C1 - Wet NEPH humidifier removed |
---|---|
Affected Datastreams: | sgpaipfitrh1ogrenC1.c1 |
Description | The humidifier was removed to fix a leak. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2012-04-07 | 00:00:00 | 2012-04-09 | 23:59:00 | Missing |
2012-04-14 | 00:00:00 | 2012-04-16 | 23:59:00 | Missing |
Subject: | SGP/AOS/C1 - Missing data |
---|---|
Affected Datastreams: | sgpaoscpcC1.a1 |
Description | Data are missing and unrecoverable. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2000-01-01 | 00:00:00 | 2016-04-01 | 00:00:00 | Suspect |
Subject: | SGP/AIP/C1 - NEPH T/RH values from thermal gradients under review |
---|---|
Affected Datastreams: | sgpaipfitrh1ogrenC1.c1 |
Description | Due to the thermal inertia of the instrument wall, slow signal response and a radial temperature gradient inside the instrument cavity, the reported RH and temperature inside the wet nephelometer may be off by up to 2-10% RH and 1-2 degrees C. The data are currently under review. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-03-01 | 00:12:38 | 2014-09-30 | 23:59:00 | Note |
Subject: | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing |
---|---|
Affected Datastreams: | sgpaosacsmC1.a1 |
Description | A closure analysis is currently being perform on the ACSM, TDMA, and wet and dry NOAA NEPHELOMETERS at SGP C1. If problems are found during this analysis, they will be promptly documented through the DQR system. The results of this analysis will be made available once completed. Please contact the respective instrument mentors with any questions regarding these data over the specified period. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
1998-12-19 | 00:00:00 | 2016-09-01 | 23:59:59 | Note |
Subject: | SGP/AIP/C1 - Reprocess: NaNs in Data |
---|---|
Affected Datastreams: | sgpaipfitrh1ogrenC1.c1 |
Description | Notice that some files during this period have NaNs in them |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2013-06-16 | 00:00:00 | 2013-06-20 | 23:59:59 | Missing |
2014-11-21 | 00:00:00 | 2014-12-31 | 23:59:59 | Missing |
2011-03-09 | 00:00:00 | 2011-03-09 | 23:59:59 | Missing |
2013-06-05 | 00:00:00 | 2013-06-05 | 23:59:59 | Missing |
2014-04-23 | 00:00:00 | 2014-04-29 | 23:59:59 | Missing |
2013-03-14 | 00:00:00 | 2013-03-17 | 23:59:59 | Missing |
2011-01-16 | 00:00:00 | 2011-01-18 | 23:59:59 | Missing |
2014-08-17 | 00:00:00 | 2014-08-21 | 23:59:59 | Missing |
2014-05-08 | 00:00:00 | 2014-05-12 | 23:59:59 | Missing |
2011-03-02 | 00:00:00 | 2011-03-02 | 23:59:59 | Missing |
2014-09-12 | 00:00:00 | 2014-09-24 | 23:59:59 | Missing |
2013-03-22 | 00:00:00 | 2013-04-14 | 23:59:59 | Missing |
2011-04-18 | 00:00:00 | 2011-04-19 | 23:59:59 | Missing |
2014-05-28 | 00:00:00 | 2014-06-03 | 23:59:59 | Missing |
Subject: | SGP/TDMA/C1 - Missing Data |
---|---|
Affected Datastreams: | sgptdmahygC1.b1 |
Description | Data is missing due to an instrument problem, and is not recoverable. |
Measurements |
|