DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D030811.1 | SGP/MFRSR/C1 - Broadband and 500nm data suspect | sgpmfrsrC1.a0 | Suspect |
D051227.5 | SGP/MFRSR/C1 - Shadowband Misalignment | sgpmfrsrC1.b1 | Incorrect |
D161007.1 | SGP/MFRSR/C1 - Shading problem | sgpmfrsrC1.a0 | Incorrect |
D170605.6 | SGP/MFRSR/C1 - Missing Irradiance Signal | sgpmfrsrC1.b1 | Incorrect |
D030531.1 | SGP/MFRSR/C1 - Shadowband misalignment | sgpmfrsrC1.a0 | Suspect |
D051123.4 | SGP/MFRSR/C1 - Missing data | sgpmfrsrC1.a0 | Missing |
D130312.5 | SGP/MFRSR/C1 - Minor shading problem | sgpmfrsrC1.a0 | Suspect |
D101220.2 | SGP/MFRSR/C1 - Shadowband Misalignment | sgpmfrsraod1michC1.c1 | Incorrect |
D130619.3 | SGP/MFRSR/C1 - Loggerboard problem | sgpmfrsrC1.a0 | Incorrect |
D151207.3 | SGP/MFRSR/C1 - Time drift | sgpmfrsrC1.b1 | Suspect |
D161128.12 | SGP/MFRSR/C1 - Possibly a poor signal cable connection | sgpmfrsrC1.b1 | Incorrect |
D110607.4 | SGP/MFRSR/C1 - Shading Problem | sgpmfrsrC1.b1 | Incorrect |
D140731.16 | SGP/MFRSR/C1 - Reprocessing Notice (RID 1206) | sgpmfrsrC1.a0 | Note |
D000610.1 | SGP/MFRSR/C1 - A/D Converter locked up | sgpmfrsrC1.a0 | Incorrect |
D170605.5 | SGP/MFRSR/C1 - Incorrect Nighttime Offset Determination | sgpmfrsrC1.b1 | Incorrect |
D160104.7 | SGP Site Power Outage Due to Winter Storm | sgpirt200msE13.a1 | Note |
D140711.4 | SGP/MFRSR/C1 - Shadowband Misalignment | sgpmfrsrC1.b1 | Incorrect |
D981112.1 | SGP/MFRSR/C1 - Irradiances too high | sgpmfrsrC1.a0 | Suspect |
D040904.5 | SGP/MFRSR/C1 - Shadowband misalignment | sgpmfrsrC1.b1 | Suspect |
D170310.16 | SGP/Central Facility - Power Outage | sgpirt25m20sC1.a0 | Note |
D170516.5 | SGP/MFRSR/C1 - Missing Data | sgpmfrsrC1.a0 | Missing |
D030907.1 | SGP/MFRSR/C1 - Shadowband misalignment | sgpmfrsrC1.a0 | Suspect |
D020821.2 | SGP/MFRSR/C1 - Out-of-band filter leakage | sgpmfrsrC1.a0 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2003-03-06 | 00:00:00 | 2003-05-01 | 23:59:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Broadband and 500nm data suspect |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | During this time period the broadband and 500nm channels intermittently reported erroneously high values due to a problem with the instruments loggerboard. The other channels were unaffected. This problem is sometimes evident as a sudden "step" shift in the values. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2005-12-22 | 16:50:00 | 2005-12-29 | 17:05:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Shadowband Misalignment |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | Instrument not shading properly. Do not use any diffuse or direct irradiance data from this instrument during this period. Maintenance on 12/29/2005 resolved the problem. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-09-11 | 16:23:00 | 2016-09-11 | 22:08:40 | Incorrect |
2016-09-04 | 16:23:00 | 2016-09-04 | 22:08:40 | Incorrect |
2016-09-03 | 16:23:00 | 2016-09-03 | 22:08:40 | Incorrect |
2016-09-08 | 16:23:00 | 2016-09-08 | 22:08:40 | Incorrect |
2016-09-09 | 16:23:00 | 2016-09-09 | 22:08:40 | Incorrect |
2016-09-10 | 16:23:00 | 2016-09-10 | 22:08:40 | Incorrect |
2016-09-17 | 16:23:00 | 2016-09-17 | 22:08:40 | Incorrect |
2016-09-15 | 16:23:00 | 2016-09-15 | 22:08:40 | Incorrect |
2016-09-14 | 16:23:00 | 2016-09-14 | 22:08:40 | Incorrect |
2016-09-13 | 16:23:00 | 2016-09-13 | 22:08:40 | Incorrect |
2016-09-19 | 16:23:00 | 2016-09-19 | 22:08:40 | Incorrect |
2016-09-06 | 16:23:00 | 2016-09-06 | 22:08:40 | Incorrect |
2016-09-16 | 16:23:00 | 2016-09-16 | 22:08:40 | Incorrect |
2016-09-20 | 16:23:00 | 2016-09-20 | 20:26:40 | Incorrect |
2016-09-07 | 16:23:00 | 2016-09-07 | 22:08:40 | Incorrect |
2016-09-01 | 16:23:00 | 2016-09-01 | 22:08:40 | Incorrect |
2016-09-02 | 16:23:00 | 2016-09-02 | 22:08:40 | Incorrect |
2016-09-12 | 16:23:00 | 2016-09-12 | 22:08:40 | Incorrect |
2016-09-05 | 16:23:00 | 2016-09-05 | 22:08:40 | Incorrect |
2016-09-18 | 16:23:00 | 2016-09-18 | 22:08:40 | Incorrect |
Subject: | SGP/MFRSR/C1 - Shading problem |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | There is a shading problem encompassing the hours 16:23 - 22:09 each day. This very likely resulted from the 90-day calibration and the head not being reinstalled in quite in the same position. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-31 | 20:42:00 | 2017-03-31 | 20:52:40 | Incorrect |
2017-03-31 | 20:58:00 | 2017-03-31 | 21:08:20 | Incorrect |
2017-04-05 | 17:47:00 | 2017-04-05 | 18:45:20 | Incorrect |
2017-04-05 | 18:52:00 | 2017-04-05 | 19:09:40 | Incorrect |
2017-04-30 | 22:48:00 | 2017-05-01 | 01:12:40 | Incorrect |
2017-05-01 | 14:40:40 | 2017-05-01 | 19:13:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Missing Irradiance Signal |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | Data values go to zero for all channels during these periods; it is most likely a connector issue. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2003-05-29 | 15:30:00 | 2003-06-02 | 16:28:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Shadowband misalignment |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | During the time period specified, the shadowband was misaligned. Portions of the data have acceptable shading and can be used by experienced data users applying appropriate screening. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2002-01-30 | 18:02:00 | 2002-02-11 | 15:02:00 | Missing |
1998-10-31 | 11:12:00 | 1998-11-03 | 14:48:00 | Missing |
2007-12-22 | 17:04:00 | 2007-12-24 | 16:04:00 | Missing |
1998-03-09 | 18:07:00 | 1998-03-11 | 17:07:00 | Missing |
2000-08-12 | 13:10:00 | 2000-08-14 | 19:11:00 | Missing |
2002-03-09 | 20:02:00 | 2002-03-12 | 07:02:00 | Missing |
2006-02-17 | 20:02:00 | 2006-02-21 | 16:03:00 | Missing |
2001-08-18 | 00:02:00 | 2001-08-20 | 16:02:00 | Missing |
2011-06-12 | 18:00:00 | 2011-06-14 | 19:00:00 | Missing |
2006-07-15 | 11:04:00 | 2006-07-17 | 14:04:00 | Missing |
1998-01-31 | 17:05:00 | 1998-02-02 | 17:07:00 | Missing |
1998-12-25 | 21:11:00 | 1998-12-28 | 15:26:00 | Missing |
2008-04-26 | 00:04:00 | 2008-04-28 | 19:04:00 | Missing |
2010-11-10 | 22:10:00 | 2010-11-12 | 16:10:00 | Missing |
1997-07-19 | 06:08:00 | 1997-07-21 | 17:06:00 | Missing |
2006-12-18 | 21:04:00 | 2006-12-20 | 21:30:00 | Missing |
2004-07-23 | 20:02:00 | 2004-07-26 | 14:55:00 | Missing |
1998-04-27 | 11:07:00 | 1998-04-29 | 20:08:00 | Missing |
1997-06-05 | 18:06:00 | 1997-06-09 | 13:22:00 | Missing |
1999-03-13 | 23:11:00 | 1999-03-15 | 19:54:00 | Missing |
Subject: | SGP/MFRSR/C1 - Missing data |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | Data are missing and unrecoverable. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2012-11-12 | 18:00:00 | 2012-12-17 | 21:00:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Minor shading problem |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | A minor shading problem occurred between hours of 1800 and 2100 UTC. The problem is very minor and the data are likely OK to use. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2010-11-23 | 19:09:00 | 2010-12-03 | 17:00:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Shadowband Misalignment |
---|---|
Affected Datastreams: | sgpmfrsraod1michC1.c1 |
Description | Shadowband slipped on the motor shaft. Maintenance resolved the issue on 20101203 @ 1700 UTC. The shading issue is particularly bad. Do not use data covered by this DQR. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-04-07 | 19:26:00 | 2011-07-18 | 15:49:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Loggerboard problem |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | There was a problem with the data logger that was in use during the period covered by this DQR. The data appear fine, but a problem was discovered when the Angstrom exponent was calculated. We do not know the root cause with the loggerboard. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-08-25 | 18:08:00 | 2015-11-05 | 12:00:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Time drift |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | A computer update caused the automatic setting of time to stop. Reports say the logger drifted as much as 4 min 41 sec. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2016-11-18 | 15:42:02 | 2016-11-18 | 18:49:20 | Incorrect |
2016-11-20 | 15:28:40 | 2016-11-20 | 21:28:20 | Incorrect |
Subject: | SGP/MFRSR/C1 - Possibly a poor signal cable connection |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | Possibly a poor signal cable connection has resulted in two periods of bad data. These data covered by this DQR are not usable. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2011-04-05 | 14:00:00 | 2011-04-05 | 19:00:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Shading Problem |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | On 20110405 in the morning there is a signature of a shading issue. This is the only time it is evident. Possible it was noticed right away during a regular site visit. Data the rest of the month look fine. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2006-07-26 | 09:24:00 | 2014-04-29 | 23:59:40 | Note |
Subject: | SGP/MFRSR/C1 - Reprocessing Notice (RID 1206) |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | See below for details of reprocessing that was completed on 20160115. Original DQR category: 3 (Yellow, Suspect). ----------------------------------------------------------- Improved code is now available to better deal with nighttime offset information. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2000-05-26 | 00:00:00 | 2000-06-09 | 23:59:00 | Incorrect |
1998-10-02 | 00:00:00 | 1998-10-14 | 23:59:00 | Incorrect |
1998-06-09 | 00:00:00 | 1998-07-07 | 20:00:00 | Incorrect |
1997-08-17 | 06:00:00 | 1997-08-20 | 16:00:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - A/D Converter locked up |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | During the time periods specified, the A/D converter malfunctioned. The data are useless. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-05-04 | 11:38:00 | 2017-05-04 | 01:16:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Incorrect Nighttime Offset Determination |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | There was a glitch in the nighttime offset calculation which affects the entire day of data. If this data is absolutely necessary contact the instrument mentor. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2015-12-26 | 00:00:00 | 2016-01-01 | 23:59:59 | Note |
Subject: | SGP Site Power Outage Due to Winter Storm |
---|---|
Affected Datastreams: | sgpirt200msE13.a1 |
Description | On Saturday and Sunday, December 26 and 27th a powerful winter storm (Goliath) brought sleet and freezing rain to much of Oklahoma with wind gusts of 45 MPH at times. The result was very very slick roads and mass power outages. The SGP central facility lost power at 8:55 PM CDT on Sunday, December 27th. The power was restored on 7:56 PM CDT on Wednesday, December 30th Site Emergency procedures were activated and Technicians drained water out of instrument reservoirs and other measures as needed, early Monday morning. On Wednesdays as roads and power started was restored to much of the areas east of I-35 , Site Technicians performed maintenance on several Extended Facilities. During this period, data are potentially missing, incorrect, or suspect. Carefully inspect the data before use. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2014-06-17 | 16:30:00 | 2014-06-17 | 18:45:00 | Incorrect |
2014-06-14 | 16:30:00 | 2014-06-14 | 18:45:00 | Incorrect |
2014-06-16 | 16:30:00 | 2014-06-16 | 18:45:00 | Incorrect |
2014-06-15 | 16:30:00 | 2014-06-15 | 18:45:00 | Incorrect |
2014-06-12 | 16:30:00 | 2014-06-12 | 18:45:00 | Incorrect |
2014-06-13 | 16:30:00 | 2014-06-13 | 18:45:00 | Incorrect |
2014-06-11 | 16:30:00 | 2014-06-11 | 18:45:00 | Incorrect |
Subject: | SGP/MFRSR/C1 - Shadowband Misalignment |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | Shading problem during the dates/times covered by this DQR. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
1998-10-14 | 00:00:00 | 1998-10-24 | 23:59:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Irradiances too high |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | After the A/D convertor lock-up was corrected on 981014, the irradiances as measured by the MFRSR were about 20% too high. 981018 is a particular egregious example of measured irradiances that are too large. Cause of the problem is not known, though suspect some moisture got into the instrument (this was a period of much precipitation). After 981024, the MFRSR settled down and the irradiances returned to a normal range. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2004-05-29 | 18:30:00 | 2004-06-03 | 12:00:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Shadowband misalignment |
---|---|
Affected Datastreams: | sgpmfrsrC1.b1 |
Description | During the time period specified, the shadowband was misaligned. Portions of the data have acceptable shading and can be used by experienced data users applying appropriate screening. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-05 | 00:23:01 | 2017-03-09 | 14:30:21 | Note |
Subject: | SGP/Central Facility - Power Outage |
---|---|
Affected Datastreams: | sgpirt25m20sC1.a0 |
Description | Automated scripts show that these instruments were affected by a power outage at the central facility resulting in missing data. Most of the instruments came back online on March 6th but some took longer. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-03-27 | 17:08:00 | 2017-03-27 | 20:08:40 | Missing |
2017-03-28 | 18:07:20 | 2017-03-28 | 19:10:20 | Missing |
2017-05-08 | 14:08:00 | 2017-05-08 | 15:08:40 | Missing |
2017-05-18 | 22:08:00 | 2017-05-18 | 23:08:40 | Missing |
2017-04-14 | 12:07:40 | 2017-04-14 | 13:08:40 | Missing |
2017-05-10 | 15:07:40 | 2017-05-10 | 16:08:40 | Missing |
Subject: | SGP/MFRSR/C1 - Missing Data |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | Missing data caused by scheduled power outages and one period by a tracker problem. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2003-09-05 | 19:00:00 | 2003-09-09 | 19:59:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Shadowband misalignment |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | During the time period specified, the shadowband was misaligned. Portions of the data have acceptable shading and can be used by experienced data users applying appropriate screening. |
Measurements |
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2001-09-19 | 00:00:00 | 2002-10-28 | 23:59:00 | Suspect |
Subject: | SGP/MFRSR/C1 - Out-of-band filter leakage |
---|---|
Affected Datastreams: | sgpmfrsrC1.a0 |
Description | Testing of the batch of centrovision filters being used in the MFRSR C1 instrument indicate that the first four filters leak 'out of band' light. The leakage for each filter is (1)415nm 3%, (2)500nm 1-2%, (3)615nm 1-2%, (4)673nm 3%. Acceptable leakage is less than 0.5%. This leakage causes an error in the irradiance values (too high) at these four wavelengths and the optical depths calculated using this data will be too high. |
Measurements |
|