Data Quality Report for Order ID: 187117
User ID: chevalierk1    Date completed: 2017-08-08
All DQRs
DQR ID Subject Affected Data Streams Data Quality Metric
D170731.1 ENA/CCN/C1 - Instrument in safe mode enaaosccn100C1.01 Incorrect
D170616.2 ENA/CPC/C1 - Data dropping out from clogged nozzle enaaoscpcfC1.b1 Suspect
D170706.1 ENA/CCN/CPC/HTDMA/NEPHELOMETER/UHSAS/C1 - Personnel on site troubleshooting enaaoscpcfC1.a1 Suspect
D170621.9 ENA/CPC/C1 - Clogged nozzle enaaoscpcfC1.b1 Missing
D170621.8 ENA/CPC/C1 - Material in the instrument nozzle enaaoscpcfC1.b1 Incorrect
 
DQR ID: D170731.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-07-28 23:00:00 2017-07-29 00:00:10 Incorrect
2017-07-29 07:00:00 2017-07-29 08:00:00 Incorrect
Subject: ENA/CCN/C1 - Instrument in safe mode
Affected Datastreams: enaaosccn100C1.01
Description Instrument went into safe mode due to low counts. Instrument was working fine, just the ambient aerosol concentration was very low.
Measurements
  • proportional_valve_voltage (CCN Proportional valve voltage)
  • CCN_T_read_TEC1 (AOS CCN observed temperature at top of column)
  • alt (Altitude above mean sea level)
  • CCN_T_read_TEC2 (AOS CCN observed temperature at middle of column)
  • CCN_Q_sample (AOS CCN volumetric flowrate of sample air)
  • qc_f_CCN (Quality check results on field: Ratio of N_CCN to concentration)
  • N_CCN_bin_number (Bin number of lowest channel of OPC include in summation of N_CCN)
  • N_CCN_fit_error (Absolute relative error between N_CCN measurement and fit)
  • supersaturation_calculated (Mean supersaturation_calculated at times to match N_CCN_mean_interquartile samples.)
  • supersaturation_calculated (Mean calculated supersaturation values for a fixed supersaturation set point.)
  • lon (East longitude)
  • first_stage_monitor_voltage (First stage monitor voltage)
  • laser_current (OPC laser current)
  • qc_N_CCN (Quality check results on field: Mean number concentration of activated condensation nuclei)
  • base_time (Base time in Epoch)
  • T_set_gradient (Temperature difference between T_set_TEC3 and T_set_TEC1)
  • N_CCN_dN (Droplet count by bin size)
  • eta (Eta factor used in calculation of supersaturation_calculated)
  • Q_sample (Volumetric flowrate of sample air)
  • concentration (Particle concentration)
  • T_sample (Temperature read of sample air entering the column)
  • T_set_TEC2 (Temperature set point at middle of column)
  • CCN_T_set_TEC2 (AOS CCN set point for temperature at middle of column)
  • supersaturation_calculated (Sample supersaturation calculated via Lance/Rose method)
  • dT_target_estimated (Estimated target thermal gradient under current operating conditions)
  • CCN_T_sample (AOS CCN temperature of sample air entering the column)
  • T_read_TEC1 (Temperature read at top of column)
  • supersaturation_setpoint (Supersaturation set point)
  • CCN_P_sample (AOS CCN sample pressure)
  • CCN_proportional_valve_voltage (CCN Proportional valve voltage)
  • aerosol_number_concentration (aerosol particle number concentration)
  • CCN_T_inlet (AOS CCN temperature of sample air at entrance to instrument)
  • size_bin (Size bins for CCN droplet count)
  • T_inlet (Temperature read of sample air at entrance to instrument)
  • seconds_after_transition (Seconds since last supersaturation set point transition)
  • laser_current (Laser current)
  • CCN_T_nafion_set (AOS CCN set temperature for the Nafion humidifier)
  • qc_N_CCN (Quality check results on field: Number concentration of CCN)
  • time_bounds (Time cell bounds)
  • eta_lookup_table (Lookup table for eta computation)
  • CCN_first_stage_monitor_voltage (First stage monitor voltage)
  • time (Time offset from midnight)
  • concentration (Computed concentration)
  • CCN_alarm_code (CCN alarm code expressed as a summation of powers of 2. 0 indicates no detected problem)
  • lat (North latitude)
  • droplet_size_bounds (Droplet bin bounds)
  • overflow (Number of particles that are larger than 10um)
  • T_read_TEC2 (Temperature read at middle of column)
  • setpoint (Supersaturation set point values)
  • T_read_gradient (Temperature difference between T_read_TEC3 and T_read_TEC1)
  • droplet_size (Size bins for CCN droplets)
  • supersaturation_calculated_target (Target supersaturation calculated via Lance/Rose method, computed from eta_target)
  • N_CCN (Mean number concentration of N_CCN)
  • CCN_T_OPC (AOS CCN temperature of the optical particle counter)
  • CCN_laser_current (AOS CCN OPC laser current)
  • temperature_std (Standard deviation of (T_read_TEC3 - T_read_TEC1))
  • setpoint_time (Supersaturation setpoint start time, relative to midnight)
  • N_CCN (Mean number concentration of activated condensation nuclei)
  • CCN_overflow (Number of particles that are larger than 10um)
  • CCN_T_nafion (AOS CCN Nafion temperature)
  • Q_sample (Volumetric flow rate of sample air)
  • T_nafion (Nafion temperature)
  • CCN_T_set_TEC3 (AOS CCN set point for temperature at bottom of column)
  • CCN_baseline_monitor (AOS CCN voltage used by DMT for OPC diagnostic purposes)
  • f_CCN (Ratio of N_CCN to concentration)
  • N_CCN (AOS number concentration of CCN)
  • CCN_temperature_gradient (Temperature difference between CCN_t_set_TEC3 and CCN_t_set_TEC1)
  • T_set_TEC3 (Temperature set point at bottom of column)
  • T_set_TEC1 (Temperature set point at top of column)
  • T_OPC (Temperature read of the optical particle counter)
  • N_CCN (Number concentration of CCN)
  • CCN_OPC_set (AOS CCN set point temperature for the optical particle counter (OPC))
  • CCN_supersaturation_set_point (AOS CCN sample supersaturation set point value reported by instrument)
  • dT_OPC (T_OPC - T_read_TEC3 temperature difference)
  • eta_target (Eta factor used in calculation of supersaturation_calculated_estimated_equilibrium_target)
  • supersaturation_set_point (Sample supersaturation set point value reported by instrument)
  • P_sample (Sample pressure)
  • T_read_TEC3 (Temperature read at bottom of column)
  • size_bin_bounds (Droplet bin bounds)
  • time_offset (Time offset from base_time)
  • CCN_temp_unstable (AOS CCN logical value, indicating whether the instrument has reached thermal gradient stabilization and is within 0.4 degrees C of the set point)
  • reported_temperature_gradient (Reported temperature difference between T_set_TEC3 and T_set_TEC1)
  • N_CCN_fit_coefs (Coefficients for quadratic fit of N_CCN vs supersaturation)
  • Q_sheath (Volumetric flowrate of sheath air)
  • N_CCN_bin_number (AOS bin number of lowest channel of OPC include in summation of N_CCN)
  • first_bin_used (First bin used in N_CCN_reduced calculation.)
  • CCN_T_read_TEC3 (AOS CCN (read) temperature at bottom of column)
  • N_CCN_fit_value (Fitted value of N_CCN versus calculated supersaturation)
  • N_CCN_reduced (N_CCN calculated excluding counts below "first_bin_used")
  • CCN_Q_sheath (AOS CCN volumetric flowrate of sheath air)
  • temp_unstable (Logical value, indicating whether the instrument has reached thermal gradient stabilization and is within 0.4 degrees C of the set point)
  • N_CCN_dN (AOS CCN droplet count by bin size)
  • CCN_T_set_TEC1 (AOS CCN set point for temperature at top of column)
  • CCN_T_inlet_set (AOS CCN set point temperature for the inlet manifold)
  • concentration (Aerosol concentration averaged over times to match N_CCN)
  • qc_N_CCN (Quality check results on field: Mean number concentration of N_CCN)
 
DQR ID: D170616.2    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-06-04 00:00:00 2017-06-05 08:09:59 Suspect
Subject: ENA/CPC/C1 - Data dropping out from clogged nozzle
Affected Datastreams: enaaoscpcfC1.b1
Description An evaluation of another CPC at the site shows a deviation between the instrument data began around this time.
Measurements
  • concentration (Particle concentration)
  • concentration (Computed concentration)
  • concentration (Aerosol concentration averaged over times to match N_CCN)
 
DQR ID: D170706.1    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-06-22 07:00:00 2017-06-30 00:00:00 Suspect
Subject: ENA/CCN/CPC/HTDMA/NEPHELOMETER/UHSAS/C1 - Personnel on site troubleshooting
Affected Datastreams: enaaoscpcfC1.a1
Description FIDO personnel on site troubleshooting and moving instruments and re-configuring sample lines. Some instruments occasionally on cabin air.
Measurements
  • proportional_valve_voltage (CCN Proportional valve voltage)
  • CCN_T_read_TEC1 (AOS CCN observed temperature at top of column)
  • CCN_T_read_TEC2 (AOS CCN observed temperature at middle of column)
  • dilution_valve_position (Dilution flow valve position)
  • CCN_Q_sample (AOS CCN volumetric flowrate of sample air)
  • sample_pressure (Sample pressure)
  • nozzle_pressure (Nozzle pressure)
  • N_CCN_bin_number (Bin number of lowest channel of OPC include in summation of N_CCN)
  • N_CCN_fit_error (Absolute relative error between N_CCN measurement and fit)
  • supersaturation_calculated (Mean supersaturation_calculated at times to match N_CCN_mean_interquartile samples.)
  • supersaturation_calculated (Mean calculated supersaturation values for a fixed supersaturation set point.)
  • first_stage_monitor_voltage (First stage monitor voltage)
  • laser_current (OPC laser current)
  • T_set_gradient (Temperature difference between T_set_TEC3 and T_set_TEC1)
  • N_CCN_dN (Droplet count by bin size)
  • eta (Eta factor used in calculation of supersaturation_calculated)
  • Q_sample (Volumetric flowrate of sample air)
  • concentration (Particle concentration)
  • T_sample (Temperature read of sample air entering the column)
  • T_set_TEC2 (Temperature set point at middle of column)
  • CCN_T_set_TEC2 (AOS CCN set point for temperature at middle of column)
  • supersaturation_calculated (Sample supersaturation calculated via Lance/Rose method)
  • unit_temperature (Unit temperature)
  • aerosol_flow_calibrated (Calibrated aerosol flow as supplied by mentor)
  • condensation_temp (Condensation temperature)
  • dT_target_estimated (Estimated target thermal gradient under current operating conditions)
  • size_distribution (Particle size distribution, limited by lower_size_limit and upper_size_limit)
  • auxiliary (Auxiliary measurement, not implemented)
  • dilution_flow_reading (Flow read)
  • purge_flow (Purge flow)
  • CCN_T_sample (AOS CCN temperature of sample air entering the column)
  • T_read_TEC1 (Temperature read at top of column)
  • supersaturation_setpoint (Supersaturation set point)
  • CCN_P_sample (AOS CCN sample pressure)
  • CCN_proportional_valve_voltage (CCN Proportional valve voltage)
  • sampling_volume (Sample volume)
  • flow (Flow measurement, not implemented)
  • sample_accumulation_time (Sample accumulation time)
  • aerosol_number_concentration (aerosol particle number concentration)
  • CCN_T_inlet (AOS CCN temperature of sample air at entrance to instrument)
  • size_bin (Size bins for CCN droplet count)
  • T_inlet (Temperature read of sample air at entrance to instrument)
  • seconds_after_transition (Seconds since last supersaturation set point transition)
  • bin_num (Bin number)
  • lower_size_limit (Lower particle size limit for each bin)
  • laser_current (Laser current)
  • temperature (Raw temperature signal)
  • upper_size_limit (Upper particle size limit for each bin)
  • CCN_T_nafion_set (AOS CCN set temperature for the Nafion humidifier)
  • ambient_pressure (Ambient pressure)
  • purge_flow_rate (Purge flow rate)
  • eta_lookup_table (Lookup table for eta computation)
  • CCN_first_stage_monitor_voltage (First stage monitor voltage)
  • raw_temperature_signal (Raw temperature signal)
  • concentration (Computed concentration)
  • CCN_alarm_code (CCN alarm code expressed as a summation of powers of 2. 0 indicates no detected problem)
  • laser_reference_voltage (Laser reference voltage)
  • cabinet_temp (Cabinet temperature)
  • droplet_size_bounds (Droplet bin bounds)
  • overflow (Number of particles that are larger than 10um)
  • T_read_TEC2 (Temperature read at middle of column)
  • droplet_size (Size bins for CCN droplets)
  • T_read_gradient (Temperature difference between T_read_TEC3 and T_read_TEC1)
  • setpoint (Supersaturation set point values)
  • supersaturation_calculated_target (Target supersaturation calculated via Lance/Rose method, computed from eta_target)
  • N_CCN (Mean number concentration of N_CCN)
  • CCN_T_OPC (AOS CCN temperature of the optical particle counter)
  • CCN_laser_current (AOS CCN OPC laser current)
  • scattering_voltage (Scatter signal)
  • temperature_std (Standard deviation of (T_read_TEC3 - T_read_TEC1))
  • setpoint_time (Supersaturation setpoint start time, relative to midnight)
  • differential_voltage (Differential_voltage, see instrument manual)
  • N_CCN (Mean number concentration of activated condensation nuclei)
  • sample_flow_rate (Sample flow rate)
  • CCN_overflow (Number of particles that are larger than 10um)
  • CCN_T_nafion (AOS CCN Nafion temperature)
  • Q_sample (Volumetric flow rate of sample air)
  • T_nafion (Nafion temperature)
  • CCN_T_set_TEC3 (AOS CCN set point for temperature at bottom of column)
  • CCN_baseline_monitor (AOS CCN voltage used by DMT for OPC diagnostic purposes)
  • f_CCN (Ratio of N_CCN to concentration)
  • reference (Reference)
  • saturation_temp (Saturation temperature)
  • sheath_flow_rate (Sheath flow rate)
  • sheath_flow (Sheath flow)
  • N_CCN (AOS number concentration of CCN)
  • CCN_temperature_gradient (Temperature difference between CCN_t_set_TEC3 and CCN_t_set_TEC1)
  • T_set_TEC3 (Temperature set point at bottom of column)
  • T_set_TEC1 (Temperature set point at top of column)
  • T_OPC (Temperature read of the optical particle counter)
  • current (Current)
  • N_CCN (Number concentration of CCN)
  • CCN_OPC_set (AOS CCN set point temperature for the optical particle counter (OPC))
  • dilution_correction_factor (Dilution correction factor)
  • CCN_supersaturation_set_point (AOS CCN sample supersaturation set point value reported by instrument)
  • optics_temp (Optics temperature)
  • dT_OPC (T_OPC - T_read_TEC3 temperature difference)
  • eta_target (Eta factor used in calculation of supersaturation_calculated_estimated_equilibrium_target)
  • supersaturation_set_point (Sample supersaturation set point value reported by instrument)
  • P_sample (Sample pressure)
  • T_read_TEC3 (Temperature read at bottom of column)
  • size_bin_bounds (Droplet bin bounds)
  • reported_temperature_gradient (Reported temperature difference between T_set_TEC3 and T_set_TEC1)
  • CCN_temp_unstable (AOS CCN logical value, indicating whether the instrument has reached thermal gradient stabilization and is within 0.4 degrees C of the set point)
  • dilution_flow_setpoint (Flow setpoint)
  • N_CCN_fit_coefs (Coefficients for quadratic fit of N_CCN vs supersaturation)
  • liquid_level (Liquid level)
  • Q_sheath (Volumetric flowrate of sheath air)
  • N_CCN_bin_number (AOS bin number of lowest channel of OPC include in summation of N_CCN)
  • first_bin_used (First bin used in N_CCN_reduced calculation.)
  • CCN_T_read_TEC3 (AOS CCN (read) temperature at bottom of column)
  • dilution_slope (Dilution slope)
  • N_CCN_fit_value (Fitted value of N_CCN versus calculated supersaturation)
  • dilution_intercept (Dilution intercept)
  • N_CCN_reduced (N_CCN calculated excluding counts below "first_bin_used")
  • orifice_pressure (Orifice pressure)
  • CCN_Q_sheath (AOS CCN volumetric flowrate of sheath air)
  • current_sensor (Current sensor)
  • fraction_full (Liquid fraction available)
  • temp_unstable (Logical value, indicating whether the instrument has reached thermal gradient stabilization and is within 0.4 degrees C of the set point)
  • N_CCN_dN (AOS CCN droplet count by bin size)
  • CCN_T_set_TEC1 (AOS CCN set point for temperature at top of column)
  • CCN_T_inlet_set (AOS CCN set point temperature for the inlet manifold)
  • concentration (Aerosol concentration averaged over times to match N_CCN)
 
DQR ID: D170621.9    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-06-14 13:30:15 2017-06-14 14:05:29 Missing
2017-06-12 14:49:40 2017-06-12 15:40:59 Missing
Subject: ENA/CPC/C1 - Clogged nozzle
Affected Datastreams: enaaoscpcfC1.b1
Description Data missing due to troubleshooting the clogged nozzle problem
Measurements
  • aerosol_flow_calibrated (Calibrated aerosol flow as supplied by mentor)
  • condensation_temp (Condensation temperature)
  • lat (North latitude)
  • alt (Altitude above mean sea level)
  • dilution_valve_position (Dilution flow valve position)
  • cabinet_temp (Cabinet temperature)
  • dilution_correction_factor (Dilution correction factor)
  • dilution_flow_reading (Flow read)
  • optics_temp (Optics temperature)
  • nozzle_pressure (Nozzle pressure)
  • time_offset (Time offset from base_time)
  • dilution_flow_setpoint (Flow setpoint)
  • liquid_level (Liquid level)
  • lon (East longitude)
  • dilution_slope (Dilution slope)
  • laser_current (OPC laser current)
  • dilution_intercept (Dilution intercept)
  • orifice_pressure (Orifice pressure)
  • laser_current (Laser current)
  • fraction_full (Liquid fraction available)
  • saturation_temp (Saturation temperature)
  • base_time (Base time in Epoch)
  • qc_concentration (Quality check results on field: Particle concentration)
  • concentration (Particle concentration)
  • ambient_pressure (Ambient pressure)
  • concentration (Aerosol concentration averaged over times to match N_CCN)
  • time (Time offset from midnight)
  • concentration (Computed concentration)
 
DQR ID: D170621.8    (Back to Top)
Start Date Start Time End Date End Time Data Quality Metric
2017-06-05 08:10:00 2017-06-14 14:25:19 Incorrect
Subject: ENA/CPC/C1 - Material in the instrument nozzle
Affected Datastreams: enaaoscpcfC1.b1
Description During this period the concentration declined steadily and then went to zero. It was later found that material was in the instrument surrounding the nozzle.
Measurements
  • concentration (Particle concentration)
  • concentration (Computed concentration)
  • concentration (Aerosol concentration averaged over times to match N_CCN)