DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D170726.7 | ENA/CO-ANALYZER/C1 - Reprocess: Valve position fields incorrect | enaaoscoC1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-06-13 | 18:00:00 | 2017-07-26 | 23:59:59 | Incorrect |
Subject: | ENA/CO-ANALYZER/C1 - Reprocess: Valve position fields incorrect |
---|---|
Affected Datastreams: | enaaoscoC1.b1 |
Description | A bug in the a1- and b1-level AOS CO ingest processes caused the valve position fields to be incorrect. This had the trickle-down effect of flagging the primary fields as incorrect via embedded QC in the b1-level product. Reprocessing is being requested now that this ingest bug has been corrected. |
Measurements |
|