DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160513.4 | SGP/KAZR/C1 - Instrument Down | sgpkazrspeccmaskgecopolC1.a0 | Missing |
D160609.5 | SGP/KAZR/C1 - Heating and cooling system failure | sgpkazrspeccmaskgecopolC1.a0 | Missing |
D160720.3 | SGP/KAZR/C1 - Missing Data | sgpkazrspeccmaskgecopolC1.a0 | Missing |
D160927.9 | SGP/MWR3C/C1 - Missing data | sgpmwr3cC1.b1 | Missing |
D161109.3 | SGP/MWR3C/C1 - Reprocess: Incorrect temperature coefficient | sgpmwr3cC1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/10/2016 | 2100 | 05/12/2016 | 1850 | Missing |
03/22/2016 | 1400 | 04/26/2016 | 1423 | Missing |
Subject: | SGP/KAZR/C1 - Instrument Down |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0 |
Description: | Instrument was down during these periods. |
Measurements: | sgpkazrspeccmaskgecopolC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/06/2016 | 1932 | 06/08/2016 | 2045 | Missing |
Subject: | SGP/KAZR/C1 - Heating and cooling system failure |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0 |
Description: | Due to a HVAC System Failure, SGP-KAZR was shut down for about 2 days. Data is missing from this time period. |
Measurements: | sgpkazrspeccmaskgecopolC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/07/2016 | 1800 | 07/12/2016 | 1659 | Missing |
Subject: | SGP/KAZR/C1 - Missing Data |
DataStreams: | sgpkazrspeccmaskgecopolC1.a0 |
Description: | Instrument was down during this period |
Measurements: | sgpkazrspeccmaskgecopolC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/13/2016 | 0000 | 08/13/2016 | 2359 | Missing |
Subject: | SGP/MWR3C/C1 - Missing data |
DataStreams: | sgpmwr3cC1.b1 |
Description: | Data are missing for the entire day |
Measurements: | sgpmwr3cC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/01/2016 | 0000 | 11/09/2016 | 1728 | Suspect |
Subject: | SGP/MWR3C/C1 - Reprocess: Incorrect temperature coefficient |
DataStreams: | sgpmwr3cC1.b1 |
Description: | The 30 GHz temperature correction coefficient has slightly changed. This results in a diurnal temperature dependence of the brightness temperature and retrievals. |
Measurements: | sgpmwr3cC1.b1:
|