DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160427.2 | AWR/HSRL/M1 - Degraded signal | awrhsrlM1.01 | Incorrect |
D160427.4 | AWR/HSRL/M1 - Hardware issue | awrhsrlM1.01 | Incorrect |
D160608.2 | AWR/HSRL/M1 - Instrument malfunction | awrhsrlM1.01 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/06/2016 | 2200 | 04/07/2016 | 2220 | Incorrect |
Subject: | AWR/HSRL/M1 - Degraded signal |
DataStreams: | awrhsrlM1.01 |
Description: | Signal degraded during this period |
Measurements: | awrhsrlM1.01:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/19/2016 | 1510 | 01/19/2016 | 2120 | Incorrect |
04/01/2016 | 1700 | 04/01/2016 | 2200 | Incorrect |
Subject: | AWR/HSRL/M1 - Hardware issue |
DataStreams: | awrhsrlM1.01 |
Description: | An unknown issue with the hardware caused artifacts in the data. |
Measurements: | awrhsrlM1.01:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/04/2016 | 0923 | 06/06/2016 | 1100 | Incorrect |
Subject: | AWR/HSRL/M1 - Instrument malfunction |
DataStreams: | awrhsrlM1.01 |
Description: | All data collected during this period is bad. Problems with the output shutter/DAQ/Seed voltage were reported by the AMF2 operator. He restarted the computer on June 4th. The system lock point was lost at that time. The lock point was manually reset from ~-9V to 2.8V on Monday June 6 around 11 UTC. |
Measurements: | awrhsrlM1.01:
|