DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140618.2 | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed | enaaosnephdryC1.b1 | Does not affect quality |
D140820.1 | ENA/NEPHELOMETER/C1 - Data processing problem | enaaosnephdryC1.b1 | Incorrect |
D141124.3 | ENA/NEPHELOMETER/C1 - Light source stopped working | enaaosnephdryC1.b1 | Incorrect |
D150506.2 | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit | enaaosnephdryC1.b1 | Incorrect |
D150701.21 | ENA/NEPHELOMETER/C1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 | enaaosnephdryC1.b1 | Does not affect quality |
D151106.2 | ENA/NEPHELOMETER/PSAP/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor stuck at 10 microns | enaaosnephdryC1.b1 | Does not affect quality |
D151112.5 | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switching | enaaosnephdryC1.b1 | Does not affect quality |
D151112.6 | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connected | enaaosnephdryC1.b1 | Does not affect quality |
D160112.1 | ENA/NEPHELOMETER/C1 - Flooded nafion cells | enaaosnephdryC1.b1 | Incorrect |
D160219.4 | ENA/NEPHELOMETER/C1 - Data not saved | enaaosnephdryC1.b1 | Missing |
D160613.1 | ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching | enaaosnephdryC1.b1 | Suspect |
D160909.13 | ENA/NEPHELOMETER/C1 - Inlet flow changed | enaaosnephdryC1.b1 | Suspect |
D161025.6 | ENA/AOS/C1 - Instrument mentors working at the site | enaaosnephdryC1.b1 | Suspect |
D161117.3 | ENA/NEPHELOMETER/C1 - Instrument offline | enaaosnephdryC1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/01/2013 | 0000 | 02/19/2014 | 1351 | Does not affect quality |
Subject: | ENA/CPC/NEPH/PSAP/C1 - Impactor not installed |
DataStreams: | enaaosnephdryC1.b1 |
Description: | No impactor installed. All data is the same as when impactor at 10 um setting. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2014 | 0351 | 03/11/2014 | 1500 | Incorrect |
Subject: | ENA/NEPHELOMETER/C1 - Data processing problem |
DataStreams: | enaaosnephdryC1.b1 |
Description: | The impactor data was not available during this period and the impactor_setting variable in these datastreams shows a constant value of 10 um. The impactor data has since been recovered. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/25/2014 | 1643 | 11/12/2014 | 1006 | Incorrect |
Subject: | ENA/NEPHELOMETER/C1 - Light source stopped working |
DataStreams: | enaaosnephdryC1.b1 |
Description: | The dry nephelometer light source stopped working. These bulbs will last only 1 to 6 year before failure. Data are incorrect and should not be used. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/22/2015 | 1700 | 04/27/2015 | 1700 | Incorrect |
Subject: | ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit |
DataStreams: | enaaosnephdryC1.b1 |
Description: | There was visit of 7 AOS instrument mentors to the ENA AOS during this period of instrument repair, calibration and modification. Instruments were occasionally off, or with no sample flow, or receiving calibration samples, etc. There was no productive sampling of ambient air during this period other than for testing. No data should be used for this period as ambient samples. Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas instrument. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2013 | 0000 | 10/01/2015 | 2359 | Does not affect quality |
Subject: | ENA/NEPHELOMETER/C1 - Reprocess: reprocessing from raw data in preparation for |
DataStreams: | enaaosnephdryC1.b1 |
Description: | End to end reprocessing to generate new a1 datastreams is needed so that when the harmonization software is released to production, we will be ready with the a1 datastreams that are input to the subsequent processing software. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/02/2015 | 1607 | 11/03/2015 | 1409 | Does not affect quality |
Subject: | ENA/NEPHELOMETER/PSAP/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor stuck at 10 microns |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Upstream impactor was stuck at the 10 micron position. Data for the period is valid but representative of an effective 10 micrometer impactor setting only. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/27/2015 | 1052 | 03/30/2015 | 1800 | Does not affect quality |
Subject: | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switching |
DataStreams: | enaaosnephdryC1.b1 |
Description: | AOS tech was working in the AOS and accidentally unplugged the impactor power supply. This left the impactor in the 10 micron state. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2014 | 1351 | 04/26/2015 | 1016 | Does not affect quality |
Subject: | ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connected |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Impactor installed and plumbed, HOWEVER, the impactor's 'size read' switch NOT connected. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/28/2015 | 1100 | 01/12/2016 | 1700 | Incorrect |
Subject: | ENA/NEPHELOMETER/C1 - Flooded nafion cells |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Nafion cells between the dry and wet nephelometer were flooded, partially blocking the air flow and possibly causing the wet nephelometer to suck in cabin air. It is unclear if/how the dry nephelometer readings were affected. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/16/2016 | 1405 | 02/16/2016 | 1718 | Missing |
Subject: | ENA/NEPHELOMETER/C1 - Data not saved |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Instrument computer rebooted for unknown reasons and got stuck on booting. No data saved for the period. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/10/2016 | 1500 | 06/12/2016 | 1100 | Suspect |
Subject: | ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Impactor power accidentally unplugged. Impactor not switching and stuck in 10um position. The quality of data variables is not affected. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/08/2016 | 0024 | 11/24/2016 | 1730 | Suspect |
Subject: | ENA/NEPHELOMETER/C1 - Inlet flow changed |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Due to a failure of another instrument on the same sample line, the sample flow for the nephelometers/humidigraph was different than specified. This potentially affected the humidigraph RH and the impactor cut point. Full effect on data quality is unknown, but is likely limited to the impactor cut point being slightly different than expected. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2016 | 0830 | 10/01/2016 | 0100 | Suspect |
Subject: | ENA/AOS/C1 - Instrument mentors working at the site |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Instrument mentors working at the site. Sample lines and electrical connections being reconfigured. Instruments that had no work done on them may still be affected. |
Measurements: | enaaosnephdryC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2016 | 1200 | 11/14/2016 | 1712 | Missing |
Subject: | ENA/NEPHELOMETER/C1 - Instrument offline |
DataStreams: | enaaosnephdryC1.b1 |
Description: | Instrument had a hard drive failure. No data collected during this period. |