DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120510.2 | SGP/QCRAD/E13 - Multiplexer failure | sgpqcrad1longE13.s2 | Incorrect |
D120510.22 | SGP/QCRAD/E13 - Noisy Tc/Td | sgpqcrad1longE13.s2 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/27/2007 | 1900 | 01/07/2008 | 1525 | Incorrect |
Subject: | SGP/QCRAD/E13 - Multiplexer failure |
DataStreams: | sgpqcrad1longE13.s2 |
Description: | The down_short_hemisp (PSP), down_long_hemisp_shaded (DIR) and up_long_hemisp (UIR) experienced frequent dropouts due to a failing multiplexer. CM replaced mux on 1/7. Don't use best estimate GSW, if "source_BestEstimate_down_short_hemisp" is -1. Documented in instrument DQR D080104.1 |
Measurements: | sgpqcrad1longE13.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/16/1999 | 0000 | 05/03/1999 | 2100 | Suspect |
Subject: | SGP/QCRAD/E13 - Noisy Tc/Td |
DataStreams: | sgpqcrad1longE13.s2 |
Description: | Case and dome temperatures for Longwave fluxes are noisy during this period corresponding to additional noise in the longwave fluxes and upwelling shortwave flux. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE13.s2:
|