DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160923.2 | OLI/PSAP/M1 - Data gaps from pump turned off | oliaospsaprM1.a0 | Incorrect |
D161211.1 | OLI/PSAP/M1 - Incorrect reset of transmittance after filter change | oliaospsaprM1.a0 | Incorrect |
D161211.2 | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, power failures | oliaospsaprM1.a0 | Incorrect |
D161211.3 | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, etc. | oliaospsaprM1.a0 | Incorrect |
D161211.4 | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off | oliaospsaprM1.a0 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/06/2016 | 1551 | 08/06/2016 | 1808 | Incorrect |
08/18/2016 | 2252 | 08/20/2016 | 2302 | Incorrect |
08/25/2016 | 1925 | 08/31/2016 | 2359 | Incorrect |
08/21/2016 | 2311 | 08/22/2016 | 1351 | Incorrect |
Subject: | OLI/PSAP/M1 - Data gaps from pump turned off |
DataStreams: | oliaospsaprM1.a0 |
Description: | Data Gaps > 1h due to Tr<0.7, pump turned off, etc. New Operator replaced previous Operator and new Operator unaware of need to change the filter. |
Measurements: | oliaospsaprM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/15/2016 | 1329 | 11/17/2016 | 1929 | Incorrect |
Subject: | OLI/PSAP/M1 - Incorrect reset of transmittance after filter change |
DataStreams: | oliaospsaprM1.a0 |
Description: | Operator MAY have changed filter but didn't reset transmission to 100%. Data invalid. |
Measurements: | oliaospsaprM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/26/2016 | 0900 | 11/26/2016 | 1022 | Incorrect |
11/11/2016 | 1124 | 11/11/2016 | 1639 | Incorrect |
11/26/2016 | 1100 | 11/26/2016 | 1208 | Incorrect |
11/19/2016 | 1209 | 11/19/2016 | 1637 | Incorrect |
11/30/2016 | 1101 | 11/30/2016 | 1622 | Incorrect |
11/27/2016 | 1813 | 11/28/2016 | 1633 | Incorrect |
11/15/2016 | 1330 | 11/17/2016 | 1929 | Incorrect |
11/01/2016 | 0908 | 11/03/2016 | 1622 | Incorrect |
11/12/2016 | 1216 | 11/12/2016 | 1652 | Incorrect |
Subject: | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, power failures |
DataStreams: | oliaospsaprM1.a0 |
Description: | The manufacturer specifies filter must be changed for transmission less than 0.7. This is not always possible with Operator staffing. This is different from D161211.1 where on that DQR, the filter was changed, but transmittance not reset. |
Measurements: | oliaospsaprM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/13/2016 | 2030 | 10/15/2016 | 1220 | Incorrect |
10/02/2016 | 1141 | 10/02/2016 | 1544 | Incorrect |
10/06/2016 | 1144 | 10/07/2016 | 0749 | Incorrect |
10/30/2016 | 0730 | 10/30/2016 | 1715 | Incorrect |
10/27/2016 | 2311 | 10/28/2016 | 1617 | Incorrect |
10/10/2016 | 0926 | 10/10/2016 | 1551 | Incorrect |
10/23/2016 | 2114 | 10/23/2016 | 2238 | Incorrect |
10/26/2016 | 0751 | 10/26/2016 | 1609 | Incorrect |
Subject: | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, etc. |
DataStreams: | oliaospsaprM1.a0 |
Description: | Manufacturer specifies transmission greater than 0.7 for valid data. This isn't always possible with Operator staffing. |
Measurements: | oliaospsaprM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/20/2016 | 1022 | 09/20/2016 | 1615 | Incorrect |
09/16/2016 | 1802 | 09/16/2016 | 2012 | Incorrect |
09/12/2016 | 2245 | 09/13/2016 | 0016 | Incorrect |
09/22/2016 | 0343 | 09/22/2016 | 1532 | Incorrect |
09/01/2016 | 0000 | 09/01/2016 | 1705 | Incorrect |
09/21/2016 | 1313 | 09/21/2016 | 1612 | Incorrect |
09/16/2016 | 0420 | 09/16/2016 | 1611 | Incorrect |
09/29/2016 | 0621 | 09/29/2016 | 1644 | Incorrect |
09/19/2016 | 2334 | 09/20/2016 | 0058 | Incorrect |
09/28/2016 | 2151 | 09/29/2016 | 0420 | Incorrect |
09/20/2016 | 0743 | 09/20/2016 | 1022 | Incorrect |
09/29/2016 | 0420 | 09/29/2016 | 0621 | Incorrect |
09/25/2016 | 1846 | 09/26/2016 | 1614 | Incorrect |
Subject: | OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off |
DataStreams: | oliaospsaprM1.a0 |
Description: | Vendor specifies transmission greater than 0.7 for valid data. This isn't possible with Operating staffing. |
Measurements: | oliaospsaprM1.a0:
|