DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D121107.2 | SGP/RWP/I10 - Instrument noise problem | sgp915rwpprecipmomI10.a0 | Does not affect quality |
D130430.1 | SGP/RWP/C1 - Intermittent bad data | sgp915rwpprecipmomC1.a0 | Incorrect |
D150420.3 | SGP/RWP/C1 - Reprocessed and Corrected: Missing Value Incorrectly Set | sgp915rwpprecipmomC1.a0 | Does not affect quality |
D150420.5 | SGP/RWP/I10 - Reprocessing Notice (RID 1322) | sgp915rwpprecipmomI10.a0 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/23/2012 | 1500 | 04/03/2012 | 0559 | Does not affect quality |
01/27/2013 | 1900 | 01/29/2013 | 0659 | Does not affect quality |
04/30/2011 | 0000 | 04/30/2011 | 2359 | Does not affect quality |
11/15/2012 | 1900 | 11/21/2012 | 0059 | Does not affect quality |
09/15/2012 | 1600 | 10/04/2012 | 2159 | Does not affect quality |
04/06/2012 | 1900 | 04/11/2012 | 0159 | Does not affect quality |
05/28/2011 | 0000 | 09/15/2011 | 0159 | Does not affect quality |
05/03/2011 | 1800 | 05/20/2011 | 0359 | Does not affect quality |
09/29/2011 | 1600 | 10/02/2011 | 2359 | Does not affect quality |
10/08/2012 | 1900 | 10/25/2012 | 2359 | Does not affect quality |
09/23/2011 | 1800 | 09/27/2011 | 0159 | Does not affect quality |
03/01/2012 | 2000 | 03/01/2012 | 2359 | Does not affect quality |
03/12/2012 | 1900 | 03/19/2012 | 0859 | Does not affect quality |
12/15/2012 | 2100 | 12/15/2012 | 2359 | Does not affect quality |
05/22/2011 | 0000 | 05/23/2011 | 2359 | Does not affect quality |
06/03/2012 | 0000 | 06/15/2012 | 0659 | Does not affect quality |
10/06/2011 | 1500 | 10/09/2011 | 0059 | Does not affect quality |
01/13/2013 | 1900 | 01/13/2013 | 2359 | Does not affect quality |
04/12/2012 | 1800 | 05/10/2012 | 2359 | Does not affect quality |
11/29/2012 | 1800 | 12/04/2012 | 2359 | Does not affect quality |
05/19/2012 | 1500 | 06/01/2012 | 0059 | Does not affect quality |
10/29/2012 | 1800 | 11/11/2012 | 0659 | Does not affect quality |
06/22/2012 | 1400 | 09/13/2012 | 0659 | Does not affect quality |
05/13/2012 | 1800 | 05/15/2012 | 0059 | Does not affect quality |
Subject: | SGP/RWP/I10 - Instrument noise problem |
DataStreams: | sgp915rwpprecipmomI10.a0 |
Description: | Instrument showing very high noise ratios on intermittent days, causing questionable data for Mode 0 noise values. Does not affect any other data |
Measurements: | sgp915rwpprecipmomI10.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/20/2012 | 0300 | 06/20/2012 | 1500 | Incorrect |
06/01/2012 | 0900 | 06/01/2012 | 1500 | Incorrect |
02/06/2012 | 0000 | 02/06/2012 | 1300 | Incorrect |
02/08/2012 | 1700 | 02/09/2012 | 1800 | Incorrect |
02/03/2012 | 1200 | 02/03/2012 | 2000 | Incorrect |
06/30/2012 | 0000 | 07/03/2012 | 1259 | Incorrect |
Subject: | SGP/RWP/C1 - Intermittent bad data |
DataStreams: | sgp915rwpprecipmomC1.a0 |
Description: | Intermittent malfunction of RWP, possibly due to poor connections. Problem did not recur after all computer boards were reseated and all electrical connections were reestablished. |
Measurements: | sgp915rwpprecipmomC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/12/2006 | 0000 | 04/08/2015 | 2359 | Does not affect quality |
Subject: | SGP/RWP/C1 - Reprocessed and Corrected: Missing Value Incorrectly Set |
DataStreams: | sgp915rwpprecipmomC1.a0 |
Description: | DATA REPROCESSED and CORRECTED: In the moment datastreams (precipmom and wind mom) from the 915 RWP, the snr variable, is not set to missing value when the SNR is invalid but the other moments have valid values. At these points the SNR has a value of 32767. The ingest was updated and the data were reprocessed to fix the issue. The data were Archived in July 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::rwp/fsite::nsa/ffac::nsa.C1 http://www.archive.arm.gov/discovery/#v/results/s/finst::rwp/fsite::sgp/ffac::sgp.C1 ORIGINAL DQR CATEGORY: Yellow( Suspect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-07 REPROCESS IDENTIFICATION NUMBER: RID 1320 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- Invalid SNR values are not being set to missing_value. |
Measurements: | sgp915rwpprecipmomC1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/25/2011 | 0000 | 04/08/2015 | 2359 | Does not affect quality |
Subject: | SGP/RWP/I10 - Reprocessing Notice (RID 1322) |
DataStreams: | sgp915rwpprecipmomI10.a0 |
Description: | See below for details of reprocessing that was completed on 20150918. Original DQR category: 3 (Yellow, Suspect). ----------------------------------------------------------- Invalid SNR values are not being set to missing_value. |
Measurements: | sgp915rwpprecipmomI10.a0:
|