DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D050321.2 | SGP/TSI/C1 - Instrument failure | sgptsiskycoverC1.b1 | Missing |
D050928.8 | SGP/TSI/C1 - Heater problem | sgptsiskycoverC1.b1 | Suspect |
D130701.1 | SGP/TSI/C1 - Data processing problem | sgptsiskycoverC1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/17/2005 | 1925 | 04/05/2005 | 1630 | Missing |
Subject: | SGP/TSI/C1 - Instrument failure |
DataStreams: | sgptsiskycoverC1.b1 |
Description: | The TSI-880 experienced a power and/or computer failure. It was replaced with a spare TSI-440 and data collection resumed 4/5/05. Note: this is a temporary replacement. |
Measurements: | sgptsiskycoverC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/01/2005 | 2135 | 10/06/2005 | 1650 | Suspect |
Subject: | SGP/TSI/C1 - Heater problem |
DataStreams: | sgptsiskycoverC1.b1 |
Description: | The TSI was converted from a model 880 to 660 which requires separate power for the mirror heater. The heater was not powered during this period so morning dew and other condensation was not quickly evaporated. |
Measurements: | sgptsiskycoverC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/02/2000 | 0000 | 10/19/2013 | 2359 | Does not affect quality |
Subject: | SGP/TSI/C1 - Data processing problem |
DataStreams: | sgptsiskycoverC1.b1 |
Description: | Values of count_opaque and count_thin are reals but should be integers (units = pixels). |
Measurements: | sgptsiskycoverC1.b1:
|