DQR ID | Subject | Data Streams Affected |
---|---|---|
D130221.1 | SGP/RLPROF/C1 - Reprocess: VAP error | sgp10rlprofbe1newsC1.c1, sgp10rlprofbe1newsC1.s1, sgp10rlprofmr1turnC1.c1 |
D130509.2 | SGP/RL/C1 - Temperature overlap correction unreliable | sgp60rlproftemp1newsC1.c0 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2010 | 0000 | 02/18/2013 | 0000 |
Subject: | SGP/RLPROF/C1 - Reprocess: VAP error |
DataStreams: | sgp10rlprofbe1newsC1.c1, sgp10rlprofbe1newsC1.s1, sgp10rlprofmr1turnC1.c1 |
Description: | Outdated and/or incorrect overlap correction files were used in original processing of the water vapor mixing ratio data (rlprofmr1turn). The overlap files have recently been updated. The VAP will be reprocessed with the updated overlap files. |
Measurements: | sgp10rlprofmr1turnC1.c1:
sgp10rlprofbe1newsC1.s1:
sgp10rlprofbe1newsC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/28/2012 | 0000 | 06/28/2012 | 2359 |
07/04/2012 | 0000 | 07/04/2012 | 2359 |
07/06/2012 | 0000 | 07/12/2012 | 2359 |
Subject: | SGP/RL/C1 - Temperature overlap correction unreliable |
DataStreams: | sgp60rlproftemp1newsC1.c0 |
Description: | The boresight alignment was not very responsive during the periods in question. This made it difficult to accurately estimate the overlap correction for the derivation of the temperture profiles. Thus, the temperature profiles below roughly 4 km AGL are suspect. |
Measurements: | sgp60rlproftemp1newsC1.c0:
|