DQR ID | Subject | Data Streams Affected |
---|---|---|
D121016.3 | PVC/IRT/M1 - Missing data | pvcirt200msM1.a1, pvcirtM1.b1 |
D121018.11 | PVC/IRT/M1 - Reprocess: Incorrect Configuration | pvcirt200msM1.a1, pvcirtM1.b1 |
D121218.5 | PVC/IRT/M1 - Instrument failure | pvcirt200msM1.a1, pvcirtM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2012 | 2300 | 09/05/2012 | 1306 |
Subject: | PVC/IRT/M1 - Missing data |
DataStreams: | pvcirt200msM1.a1, pvcirtM1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | pvcirt200msM1.a1:
pvcirtM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/29/2012 | 2240 | 10/18/2012 | 1259 |
Subject: | PVC/IRT/M1 - Reprocess: Incorrect Configuration |
DataStreams: | pvcirt200msM1.a1, pvcirtM1.b1 |
Description: | The SKYRAD IRT was installed temporarily (waiting for the new enclosure) in a vertical orientation, without the use of the gold mirror. When irt_comm was started, the emissivity parameter in the startup file was not changed to 1.000 compensate for the absence of the mirror, resulting in a positive bias of about 5 K. These data may be corrected by adjusting the radiance values for the difference in emmissivity. |
Measurements: | pvcirt200msM1.a1:
pvcirtM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2012 | 1420 | 12/04/2012 | 1434 |
Subject: | PVC/IRT/M1 - Instrument failure |
DataStreams: | pvcirt200msM1.a1, pvcirtM1.b1 |
Description: | The IRT failed because of water intrusion. It was replaced with a unit installed inside a ventilated enclosure. |
Measurements: | pvcirt200msM1.a1:
pvcirtM1.b1:
|