Subject: | NSA/KAZR/C1 - Reprocess: Incorrect calibration constants |
DataStreams: | nsakazrauxC1.a1, nsakazrblC1.b1, nsakazrgeC1.b1
|
Description: | The calibration constants for the modes were incorrect. The coefficients used were
BL: -34.70 dB
GE: -18.69 dB
The correct coefficients are
BL: -22.21dB
GE: -13.14 dB |
Measurements: | nsakazrblC1.b1: - reflectivity_xpol
- cal_constant_copol
- Reflectivity, copolar(reflectivity_copol)
- cal_constant_xpol
nsakazrauxC1.a1: - cal_constant_ch1_copol
- cal_constant_ch1_xpol
- cal_constant_ch2_xpol
- cal_constant_ch2_copol
nsakazrgeC1.b1: - cal_constant_xpol
- reflectivity_xpol
- Reflectivity, copolar(reflectivity_copol)
- cal_constant_copol
|