DQR ID | Subject | Data Streams Affected |
---|---|---|
D110221.4 | SBS/MWR/M1 - Missing data | sbsmwrtipM1.a1, sbsmwrlosM1.b1 |
D110221.5 | SBS/MWR/M1 - Instrument not well calibrated | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
D110221.6 | SBS/MWR/M1 - Reprocessed: Incorrect retrievals | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
D110317.2 | SBS/MWR/M1 - Temperature instability | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
D110414.3 | SBS/MWR/M1 - Instrument problem | sbsmwrM1.00, sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/15/2010 | 0000 | 01/12/2011 | 0000 |
03/11/2011 | 2000 | 03/14/2011 | 1400 |
03/17/2011 | 1500 | 03/23/2011 | 2000 |
03/27/2011 | 1300 | 04/01/2011 | 1300 |
Subject: | SBS/MWR/M1 - Missing data |
DataStreams: | sbsmwrtipM1.a1, sbsmwrlosM1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sbsmwrlosM1.b1:
sbsmwrtipM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2011 | 0000 | 01/14/2011 | 0000 |
Subject: | SBS/MWR/M1 - Instrument not well calibrated |
DataStreams: | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
Description: | After being installed the instrument needed some time to stabilize and collect some tip data. |
Measurements: | sbsmwrlosM1.b1:
sbsmwrtipM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2010 | 0000 | 04/25/2011 | 2359 |
Subject: | SBS/MWR/M1 - Reprocessed: Incorrect retrievals |
DataStreams: | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
Description: | The PWV and LWP retrievals at this site were incorrect for lack of radiosonde data. At the end of the deployment, the mentor provided new retrieval coefficients and the MWRLOS data were reprocessed and archived. The MWRTIP data have not yet been reprocessed. The correction of these data are in the reprocessing queue at a low priority. |
Measurements: | sbsmwrlosM1.b1:
sbsmwrtipM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2011 | 0000 | 02/03/2011 | 0000 |
Subject: | SBS/MWR/M1 - Temperature instability |
DataStreams: | sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
Description: | The radiometer was thermally unstable due to the extremely cold outside temperature. Data are suspect. |
Measurements: | sbsmwrlosM1.b1:
sbsmwrtipM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/14/2011 | 1400 | 03/17/2011 | 1500 |
03/23/2011 | 2000 | 03/27/2011 | 1300 |
Subject: | SBS/MWR/M1 - Instrument problem |
DataStreams: | sbsmwrM1.00, sbsmwrlosM1.b1, sbsmwrtipM1.a1 |
Description: | The fan in the dewblower malfunctioned from 3/11 to 3/30. During this time it is likely that snow deposited on the radome during snowfall events and it stayed on the radome longer than usual. |
Measurements: | sbsmwrM1.00:
sbsmwrlosM1.b1:
sbsmwrtipM1.a1:
|