DQR ID | Subject | Data Streams Affected |
---|---|---|
D031007.3 | TWP/SKYRAD/C2 - Incorrect Calibration Coefficients Applied | twpskyrad20sC2.a1, twpskyrad60sC2.a1 |
D040812.2 | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
D040812.3 | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC2.a1, twpskyrad60sC2.b1 |
D040812.5 | TWP/SKYRAD/C3 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC3.a1, twpskyrad60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/20/2002 | 2347 | 09/20/2003 | 2333 |
Subject: | TWP/SKYRAD/C2 - Incorrect Calibration Coefficients Applied |
DataStreams: | twpskyrad20sC2.a1, twpskyrad60sC2.a1 |
Description: | Incorrect calibration coefficients were entered for three instruments (PSPG, PIRD and PIRG) at the RESET last Sept. |
Measurements: | twpskyrad60sC2.a1:
twpskyrad20sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1996 | 1934 | 08/01/2004 | 2359 |
Subject: | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC1.b1:
twpskyrad20sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/04/2003 | 0000 | 07/05/2004 | 2359 |
Subject: | TWP/SKYRAD/C2 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC2.a1, twpskyrad60sC2.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad20sC2.a1:
twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2003 | 0000 | 05/19/2004 | 2200 |
Subject: | TWP/SKYRAD/C3 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC3.a1, twpskyrad60sC3.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC3.b1:
twpskyrad20sC3.a1:
|