DQR ID | Subject | Data Streams Affected |
---|---|---|
D040812.2 | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1996 | 1934 | 08/01/2004 | 2359 |
Subject: | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC1.b1:
twpskyrad20sC1.a1:
|