DQR ID | Subject | Data Streams Affected |
---|---|---|
D030822.7 | NSA/MWR/C1 - min/max/delta values incorrect | nsamwrlosC1.b1 |
D060420.8 | NSA/MWR/C1 - software upgrade (version 3.28) | nsamwrlosC1.a1, nsamwrlosC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/02/1998 | 0000 | 02/08/2003 | 2359 |
Subject: | NSA/MWR/C1 - min/max/delta values incorrect |
DataStreams: | nsamwrlosC1.b1 |
Description: | The values of valid_min, valid_max, and valid_delta for fields tkxc and tknd were incorrect. They should be 303, 333, and 0.5 K, respectively. |
Measurements: | nsamwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/02/1998 | 0000 | 02/03/1999 | 1200 |
Subject: | NSA/MWR/C1 - software upgrade (version 3.28) |
DataStreams: | nsamwrlosC1.a1, nsamwrlosC1.b1 |
Description: | The MWR operating program was upgraded to version 3.28 on 3 February 1999. This version includes a beam width correction as well as providing the capability to automatically level the elevation mirror (that is, to automatically detect and correct offsets in the elevation angle stepper motor position.) The improvement in the quality of the tip curves resulting from the auto- leveling has been dramatic: differences in the brightness temperatures at 3 airmasses (19.5 and 160.5 degrees) have been reduced from +/- 5 K to +/- 0.5 K. In order to take full advantage of this improvement to detect and reject cloudy tip curves, the minimum value of the goodness-of-fit coefficient for a valid tip curve has been increased from 0.995 to 0.998. |
Measurements: | nsamwrlosC1.a1:
nsamwrlosC1.b1:
|