DQR ID | Subject | Data Streams Affected |
---|---|---|
D000308.1 | SGP/SONDE - Questionable absolute pressures | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000308.2 | SGP/SONDE - Questionable absolute pressures | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000319.1 | SGP/SONDE - Data missing from sounding | sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1 |
D000319.2 | SGP/SONDE - Data missing from sounding | sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1 |
D990124.2 | SGP/SONDE - Dry bias in sonde RH | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2000 | 2328 | 03/04/2000 | 0051 |
Subject: | SGP/SONDE - Questionable absolute pressures |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The sonde surface pressure reading was considerably higher than the corresponding THWAPS pressure (988.0 hPa vs 982.3 hPa). The sonde readings are consistent during the sounding, however, suggesting that this is not a case of a pressure jump at launch but one of sensor offset. This problem will only affect those who use the absolute pressure values. The file altitudes, which are based on the relative pressure changes from the surface, should be fine. |
Measurements: | sgpsondewrpnC1.00:
sgpsondewrpnC1.a1:
sgpsondewrpnC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/04/2000 | 0229 | 03/04/2000 | 0359 |
Subject: | SGP/SONDE - Questionable absolute pressures |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The sonde surface pressure reading was considerably higher than the corresponding THWAPS pressure (992.9 hPa vs 983.0 hPa). The sonde readings are consistent during the sounding, however, suggesting that this is not a case of a pressure jump at launch but one of sensor offset. This problem will only affect those who use the absolute pressure values. The file altitudes, which are based on the relative pressure changes from the surface, should be fine. |
Measurements: | sgpsondewrpnC1.00:
sgpsondewrpnC1.a1:
sgpsondewrpnC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2000 | 2042 | 03/16/2000 | 2143 |
Subject: | SGP/SONDE - Data missing from sounding |
DataStreams: | sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1 |
Description: | It appears that the collection of the raw data file was interupted by the SDS. The following message appears in the raw file (sonde-hills10003161901.rawraw): 12 0 4.8 4384 589.3 -11.0 87 -12.7 274 12.8 110001110110 #Collection process shut down by SIGTERM Another raw file (sonde-hills10003162145.raw) follows with: ~LL? 73 30 4.7 22154 37.9 -63.4 1 -93.3 250 13.2 110001110110 I don't understand the cause of this problem, but am confident that it is not instrument related. These files were merged on ingest. |
Measurements: | sgpsondewrpnB1.00:
sgpsondewrpnB1.a1:
sgpsondewrpnB1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/2000 | 2043 | 03/16/2000 | 2144 |
Subject: | SGP/SONDE - Data missing from sounding |
DataStreams: | sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1 |
Description: | It appears that the SDS interupted collection of the data from this sounding. There is a large gap and the message 13 0 4.0 4116 610.8 -8.1 78 -11.3 277 8.6 010010100101 #Collection process shut down by SIGTERM 74 38 4.6 19907 54.8 -62.6 1 -92.7 /// //// 011110100101 appears in the ingested data file. This is the same type of thing that happened at B1 at the same time. See DQR D000319.1 |
Measurements: | sgpsondewrpnB6.00:
sgpsondewrpnB6.a0:
sgpsondewrpnB6.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/1992 | 0000 | 05/31/2002 | 2359 |
Subject: | SGP/SONDE - Dry bias in sonde RH |
DataStreams: | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Description: | Vaisala has confirmed ARM findings of an apparent dry bias in the relative humidity measured by RS-80H radiosondes. The cause of the dry bias is thought to be contamination of the humidity sensor by volatile organic substances originating from some plastic parts of the radiosonde. The amount of contamination is a function of the time between the date of sonde manufacture and its use. All RS-80H sondes manufactured before week 34 of 1998 will show this bias. After week 34 of 1998 Vaisala changed its packaging to reduce, but not eliminate the contamination problem. Starting with RS-80 radiosonde manufactured in late June 2000 Vaisala enclosed the sensor boom in an inert plastic shield, thereby eliminating the contamination that caused the dry bias. Starting in May 2001 at the SGP, May 2002 at the TWP, and later 2002 at the NSA, ARM has moved to using RS-90 radiosondes. These sondes are not subject to the contaminatino that caused the dry bias. Vaisala is in the process of developing an algorithm that can be used to estimate the correct RH from knowledge of the sonde age. All of the ARM sounding data have sufficient metadata available to apply the correction. Additionally, ARM has funded a Science Team effort (Milosevich) to develop a 'best' correction algorithm for the RS-80 radiosonde humidity data. When completed this algorithm will allow us to reprocess the accumulated RS-80 data and produce a new data platform with what we hope will be more accurate data. |
Measurements: | sgpsondeptuC1.00:
sgpsondeB4.a1:
sgpsondeB4.a0:
DsgpsondeptuB4.00:
sgpsondewrprB1.00:
DsgpsondeC1.a1:
DsgpsondeB5.a0:
DsgpsondeC1.a0:
sgpsondewrpnC1.a1:
sgpsondewnprC1.a1:
DsgpsondenogcptucalcC1.c1:
sgpsondewrpnB5.00:
sgpsondewrpnB4.a0:
sgpsondewnpnB1.a0:
sgpsondewnpnB1.a1:
DsgpsondeB5.a1:
sgpsondewrprB4.a1:
sgpsondewrpnB4.a1:
DsgpsondeB4.a0:
sgpsondewrpnB6.00:
DsgpsondeB4.a1:
sgpsondewrprB4.a0:
DsgpsondeptuB6.00:
sgpsondewnpnC1.a0:
sgpsondewrpnB6.a0:
DsgpsondeptuB5.00:
sgpsondewnpnC1.a1:
DsgpsondenogcwrpnB4.c1:
DsgpsondeB4.00:
sgpsondewrpnB6.a1:
DsgpsondenogcptucalcB4.c1:
DsgpsondeC1.00:
DsgpsondeptuC1.00:
DsgpsondeptucalcB1.c1:
DsgpsondeptucalcB5.c1:
sgpsondewrpnB1.00:
sgpsondewrpnB1.a1:
sgpsondewrpnB1.a0:
DsgpsondeptuB1.00:
DsgpsondeB1.00:
DsgpsondeptucalcB4.c1:
DsgpsondenogcwrpnB5.c1:
sgpsondewrpnB5.a0:
sgpsondewrpnB5.a1:
DsgpsondenogcptucalcB1.c1:
sgpsondeC1.00:
sgpsondewrprB4.00:
sgpsondeB4.00:
sgpsondewrprB1.a0:
sgpsondewrprC1.00:
sgpsondewrprB1.a1:
sgpsondewnpnB5.a0:
sgpsondeB1.a0:
DsgpsondeB5.00:
sgpsondewnpnB5.a1:
sgpsondeB1.a1:
sgpsondewnpnB4.a1:
sgpsondewnpnB6.a0:
sgpsondewnpnB6.a1:
sgpsondewnpnB4.a0:
DsgpsondenogcwrpnB1.c1:
sgpsondeB5.00:
sgpsondeB6.00:
DsgpsondenogcptucalcB5.c1:
sgpsondewrprB5.00:
DsgpsondeptucalcC1.c1:
DsgpsondenogcwrpnC1.c1:
sgpsondeC1.a1:
sgpsondeC1.a0:
DsgpsondeB1.a0:
sgpsondewrpnC1.00:
sgpsondeB5.a0:
sgpsondewrpnB4.00:
sgpsondeB5.a1:
DsgpsondeB1.a1:
sgpsondewrprC1.a1:
sgpsondewrprB5.a0:
sgpsondewrprB5.a1:
sgpsondewrpnC1.a0:
sgpsondewnprC1.a0:
sgpsondeB1.00:
sgpsondewrprC1.a0:
|