DQR ID | Subject | Data Streams Affected |
---|---|---|
D040421.3 | NSA/SKYRAD/C1 - itermittant IRT bias | nsaskyrad60sC1.b1 |
D050219.2 | TWP/SKYRAD/C2 - IRT bias | twpskyrad60sC2.b1 |
D050504.2 | TWP/GNDRAD/C2 - IRT connector | twpgndrad60sC2.b1 |
D060630.34 | TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error | twpgndrad60sC1.b1 |
D060630.35 | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error | twpgndrad60sC2.b1 |
D060630.36 | TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error | twpgndrad60sC3.b1 |
D060630.39 | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error | nsagndrad60sC1.b1 |
D060630.40 | NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error | nsagndrad60sC2.b1 |
D060630.44 | PYE/GNDRAD/M1 - Reprocessed: Longwave calibration error | pyegndrad60sM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2003 | 0200 | 06/21/2006 | 0215 |
Subject: | NSA/SKYRAD/C1 - itermittant IRT bias |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky temperature measurement then a sudden correction, as compared to AERI, appears to have been caused by a combination of a failing IR detector, a poor signal cable, a contaminated (with condensation or frost) mirror or lens, and low ambient temperatures. The positive bias was apparently not consistant in either frequency or duration but on days when the unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. The instrument was repaired and recalibrated and the signal cable replaced. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2005 | 0454 | 04/17/2005 | 2023 |
Subject: | TWP/SKYRAD/C2 - IRT bias |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The IRT was measuring higher sky temperatures than AERI. The problem was corrected by changing the mirror. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2005 | 2330 | 04/03/2005 | 0240 |
Subject: | TWP/GNDRAD/C2 - IRT connector |
DataStreams: | twpgndrad60sC2.b1 |
Description: | The IRT had a problem with the connector of either the signal or power cable which produced an output of 0 volts equivalent to a measurement of 223K. The problem was corrected when the connectors were cleaned and tightened. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2004 | 0700 | 12/21/2005 | 2204 |
Subject: | TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 2202 | 12/21/2005 | 2149 |
Subject: | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2004 | 0100 | 12/21/2005 | 2137 |
Subject: | TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2004 | 1600 | 12/21/2005 | 1855 |
Subject: | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/17/2004 | 0037 | 12/22/2005 | 2152 |
Subject: | NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2005 | 0000 | 09/14/2005 | 0000 |
Subject: | PYE/GNDRAD/M1 - Reprocessed: Longwave calibration error |
DataStreams: | pyegndrad60sM1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in February 2007. |
Measurements: | pyegndrad60sM1.b1:
|