DQR ID | Subject | Data Streams Affected |
---|
D950109.1 | Questionable surface value 1/2/95 11:32 | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950109.2 | Questionable surface value 1/3/95 11:29 | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950304.1 | Reprocess: Sounding 950222:0530 done in research mode | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950310.1 | Bad surface values in sounding 3/9/95:0530 | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950316.1 | Reprocess: Sounding 950315:1429 done in research mode | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950417.1 | Reprocess: Sounding 950413:1130 done in research mode | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950419.1 | Bad surface values in sounding 4/17/95 14:29 | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D950816.5 | bad surface tdry value in sounding 17:55 | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.1 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.2 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.3 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.4 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.5 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D951006.6 | Apparent Interference Signal in Sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D960129.1 | Bad surface pressure sounding | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D960229.2 | SGP/SONDE/C1 - incorrect factory RH calib-radiosondes | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Subject: | Questionable surface value 1/2/95 11:32 |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sgpsondec1wrpn
Subject: Questionable surface value 1/2/95 11:32
Date Submitted: 1/9/95
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: sgpsondec1
What level data: (raw,a0,a1,b1,c1 etc): all
What location was the data collected at: cf
Period of time in question
Begin Date 1/02/95 Time 11:32:00 (GMT)
End Date 1/02/95 Time 11:32:02 (GMT)
Data should be labeled:
_x_ questionable ___ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
It appears that incorrect surface values of temperature and relative
humidity were entered for the subject sounding.
Other observations/measurements impacted by this problem:
None
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Questionable surface value 1/3/95 11:29 |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sgpsondec1wrpn
Subject: Questionable surface value 1/3/95 11:29
Date Submitted: 1/9/95
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: sgpsondec1
What level data: (raw,a0,a1,b1,c1 etc): all
What location was the data collected at: cf
Period of time in question
Begin Date 1/03/95 Time 11:29:00 (GMT)
End Date 1/03/95 Time 11:29:02 (GMT)
Data should be labeled:
_x_ questionable ___ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
It appears that incorrect surface values of temperature and relative
humidity were entered for the subject sounding.
Other observations/measurements impacted by this problem:
None
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Reprocess: Sounding 950222:0530 done in research mode |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform:
Subject: Sounding 950222:0530 Done in Research Mode
Date Submitted: 3/4/95
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: SGPSONDEC1
What level data: (raw,a0,a1,b1,c1 etc): ALL
What location was the data collected at: SGP/CF
Period of time in question
Begin Date 2/22/95 Time 05:30 (GMT)
End Date 2/22/95 Time 07:30 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
_X_ others
Discussion of Problem: This sounding was done in RESEARCH mode for
both thermodynamic as well as wind data. The
normal mode is to for RESEARCH winds only. As
a result, both the ascending and descending
flights of the sonde are included in the data
file and none of the data are smoothed or
filtered.
Other observations/measurements impacted by this problem: I think that
some of the radiation codes that use the sonde
data as input don't handle the descending samples
well. I'm not sure which ones may be affected.
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.) Users should be notified and,
if (and only if) requested, the sounding could be reprocessed either using
the software developed at the EC or the digi-CORA.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Bad surface values in sounding 3/9/95:0530 |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sgpsondeC1
Subject: Bad surface values in sounding 3/9/95:0530
Date Submitted: 3/10/95
Submitted By: _x_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): all
What location was the data collected at: sgp/cf
Period of time in question
Begin Date 3/09/95 Time 05:30 (GMT)
End Date 3/09/95 Time 05:30 (GMT)
Data should be labeled:
___ questionable ___ All data fields affected
_x_ incorrect _x_ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears that the operator entered incorrect
values for the surface temperature and humidity for this sounding. The
entered values are 24.8 degC (temp) and 19 %RH (humidity) which are most
likely the values obtained from the met instruments inside the trailer
rather from the sonde on the launch platfrom. The next sample values of
-3.1 degC (temp) and 81 %RH (humidity) appear to be valid.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Reprocess: Sounding 950315:1429 done in research mode |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | P950316.1 Barry Lesht
SGP/SONDE.C1.950315.950315 - Sounding 950315:1429 Done in Research Mode
DQR No: Platform:
Subject: Sounding 950315:1429 Done in Research Mode
Date Submitted: 3/4/95
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: SGPSONDEC1
What level data: (raw,a0,a1,b1,c1 etc): ALL
What location was the data collected at: SGP/CF
Period of time in question
Begin Date 3/15/95 Time 14:29 (GMT)
End Date 3/15/95 Time 16:29 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
_X_ others
Discussion of Problem: This sounding was done in RESEARCH mode for
both thermodynamic as well as wind data. The
normal mode is to for RESEARCH winds only. As
a result, both the ascending and descending
flights of the sonde are included in the data
file and none of the data are smoothed or
filtered.
Other observations/measurements impacted by this problem: I think that
some of the radiation codes that use the sonde
data as input don't handle the descending samples
well. I'm not sure which ones may be affected.
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.) Users should be notified and,
if (and only if) requested, the sounding could be reprocessed either using
the software developed at the EC or the digi-CORA.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Reprocess: Sounding 950413:1130 done in research mode |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform:
Subject: Sounding 950413:1130 Done in Research Mode
Date Submitted: 4/17/95
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: SGPSONDEC1
What level data: (raw,a0,a1,b1,c1 etc): ALL
What location was the data collected at: SGP/CF
Period of time in question
Begin Date 4/13/95 Time 11:30 (GMT)
End Date 4/13/95 Time 13:30 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
___ incorrect ___ Only some data fields affected
___ wrong calibration
_X_ others
Discussion of Problem: This sounding was done in RESEARCH mode for
both thermodynamic as well as wind data. The
normal mode is to for RESEARCH winds only. As
a result, both the ascending and descending
flights of the sonde are included in the data
file and none of the data are smoothed or
filtered.
Other observations/measurements impacted by this problem: I think that
some of the radiation codes that use the sonde
data as input don't handle the descending samples
well. I'm not sure which ones may be affected.
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.) Users should be notified and,
if (and only if) requested, the sounding could be reprocessed either using
the software developed at the EC or the digi-CORA.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Bad surface values in sounding 4/17/95 14:29 |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: SGPSONDEC1
Subject: Bad surface values in sounding 4/17/95 14:29
Date Submitted: 4/19/95
Submitted By: Barry Lesht _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: All
What level data: (raw,a0,a1,b1,c1 etc):
What location was the data collected at: SGP/CF
Period of time in question
Begin Date 4/17/95 Time 14:29 (GMT)
End Date 4/17/95 Time 15:51 (GMT)
Data should be labeled:
___ questionable ___ All data fields affected
_X_ incorrect _X_ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
The operator entered incorrect surface values for both RH and pressure.
Because of the way altitude is calculated, the incorrect pressure value
has the effect of making all subsequent altitude values incorrect. The
RH error only affects the first sample.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | bad surface tdry value in sounding 17:55 |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: SGPSONDEC1
Subject: Bad surface tdry value in sounding 8/11/95 17:55
Date Submitted: 8/16/95
Submitted By: Barry Lesht _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: All
What level data: (raw,a0,a1,b1,c1 etc):
What location was the data collected at: SGP/CF
Period of time in question
Begin Date 8/11/95 Time 17:55 (GMT)
End Date 8/11/95 Time 17:55 (GMT)
Data should be labeled:
___ questionable ___ All data fields affected
_X_ incorrect _X_ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
An incorrect value was entered by the operator for surface temperature.
The value entered as 28.8 should be 31.8. This error was noted in the
site operations log by the operator shortly after it occurred.
The remaining values in the sounding should be fine.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951002:2330
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/03/95 Time XX:XX:XX (GMT)
End Date 10/03/95 Time 00:54:12 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 14.5 km. At this point the RH signal
drops out, and both the pressure and temperature signals are noisy and
look unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951003:0830
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/03/95 Time XX:XX:XX (GMT)
End Date 10/03/95 Time 09:40:20 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 14.0 km. At this point the RH signal
drops out, and both the pressure and temperature signals are noisy and
look unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951003:1429
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/03/95 Time XX:XX:XX (GMT)
End Date 10/03/95 Time 15:25:24 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 12.7 km. At this point the RH signal
drops out, and both the pressure and temperature signals are noisy and
look unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951004:0530
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/04/95 Time XX:XX:XX (GMT)
End Date 10/04/95 Time 07:07:50 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 12.0 km. At this point the RH signal
drops out, and both the pressure and temperature signals are noisy and
look unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951004:0834
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/04/95 Time XX:XX:XX (GMT)
End Date 10/04/95 Time 09:45:30 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 11.5 km. At this point the RH signal
and the pressure and temperature signals are noisy and look unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Apparent Interference Signal in Sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sondeC1
Subject: Apparent Interference Signal in Sounding CF:951004:1130
Date Submitted: October 6, 1995
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement:
What level data: (raw,a0,a1,b1,c1 etc): All
What location was the data collected at: CF
Period of time in question
Begin Date 10/04/95 Time XX:XX:XX (GMT)
End Date 10/04/95 Time 13:12:02 (GMT)
Data should be labeled:
___ questionable _X_ All data fields affected
_X_ incorrect ___ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem: It appears as if the ground station started to
receive data from a foreign sonde (i.e., not the sonde originally being
tracked) at an altitude of about 17.0 km. At this point the RH signal
drops out and the pressure and temperature signals are noisy and look
unnatural.
Other observations/measurements impacted by this problem:
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | Bad surface pressure sounding |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | DQR No: Platform: sgpsondeC1
Subject: Bad surface pressure sounding 12/13/95:0531
Date Submitted: 1/29/96
Submitted By: _X_ Instrument Mentor
___ EST Member
___ Science Team Member
___ Other _____________________________
For questions or problems, please contact the ARM Experiment Center at
509-375-6898 or via email at dqr@arm.gov.
Platform/Measurement: sgpsondewrpnC1/all
What level data: (raw,a0,a1,b1,c1 etc):
What location was the data collected at: CF
Period of time in question
Begin Date 12/13/95 Time 05:31 (GMT)
End Date Time (GMT)
Data should be labeled:
___ questionable ___ All data fields affected
_X_ incorrect _X_ Only some data fields affected
___ wrong calibration
___ others
Discussion of Problem:
The surface pressure for this sounding was entered by the operator as 1010 hPa.
The first pressure value reported by the sonde is 971.8 hPa. Pressure at SMOS13
at 05:30 was 973.7 hPa. The reason for the discrepency between the apparent
surface pressure and that entered by the operator is unknown.
Other observations/measurements impacted by this problem:
None
Suggested Corrections of the Problem: (e.g. change calibration factor and
recompute, flag data with this comment, etc.)
Flag data.
Data Processing Notes Date |
Measurements: | sgpsondewrpnC1.a1: - Retrieved pressure profile(pres)
- Time offset of tweaks from base_time(time_offset)
- lat(lat)
- Relative humidity inside the instrument enclosure(rh)
- V-component(v_wind)
- Ascent Rate(asc)
- U-component(u_wind)
- Dummy altitude for Zeb(alt)
- Dry bulb temperature(tdry)
- Surface dew point temperature(dp)
- Mean Wind Speed(wspd)
- lon(lon)
- Wind Direction(deg)
- Wind Status(wstat)
- base time(base_time)
sgpsondewrpnC1.a0: - base time(base_time)
- Dummy altitude for Zeb(alt)
- Retrieved pressure profile(pres)
- Ascent Rate(asc)
- Wind Direction(deg)
- Surface dew point temperature(dp)
- lat(lat)
- Dry bulb temperature(tdry)
- Wind Status(wstat)
- Relative humidity inside the instrument enclosure(rh)
- Mean Wind Speed(wspd)
- lon(lon)
- Time offset of tweaks from base_time(time_offset)
|
Subject: | SGP/SONDE/C1 - incorrect factory RH calib-radiosondes |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1
|
Description: | Ongoing investigation of differences between preciptiable water vapor
measured by the microwave radiometer and precipitable water vapor estimated
by integrating the radiosonde-measured profile of water vapor density, along
with isolated periods of anomalous behavior in the AERI/LBLRTM QME results
led to the suggestion that the calibration of some radiosondes may be suspect.
One group of radiosondes, calibrated in November 1994, seemed particularly
irregular. I met with the manufacturer on February 7 to show them my
results and discuss this issue with them. The manufacturer agreed that
the data were suggestive of a calibration problem and offered to review
their calibration records for the period in question.
The initial results of the manufacture's investigation were sent to me
on February 27. The manufacturer found that radiosondes assembled
between 10/26/94 and 11/22/94 were incorrectly calibrated. The incorrect
calibration affected only the relative humidity measurement; the nature
of the calibration error is non-linear, with the radiosondes meeting
the specified +/- 2% RH accuracy at both 20%RH (+45 degC) and 0%RH (room
temperature). The error in calibration occurs at high RH and low temperature
(approximate saturation at -30 degC) and can be as large as 10%RH with
the radiosonde reading high. The direction of this error agrees with
the observation of negative integrated residuals in the AERI ch 1 region
made by Tony Clough and the tendency toward greater than 'normal' differences
in PWV when compared to the MWR when these sondes were used.
The cause of the calibration error was the addition of new, high RH,
quality control check (-30 degC) into the manufacturing stream that
required a change in the calibration software. The new software mistakenly
loaded an incorrect parameter file that caused the calibration to be
off at high RH - this was not noticed at the time because the radiosondes
passed the regular production QC checks and the failure to agree at the
new QC value was thought to result from a problem with the new reference
standard. Troubleshooting the entire calibration system in an effort
to incorporate the new QC check was conducted along with normal production,
and took about a month to complete. The troubleshooting including adjustment
of the calibration software, at which time the correct parameter file
happened to be reloaded, and after which the radiosondes met the desired
specification at the high RH value.
According to the manufacturer, approximately 1900 radiosondes were calibrated
during the period in question. Sondes from this period were used at all
CART facilities. At least 1178 soundings involving these sondes were
included in the QME results I looked at (some QME results were discarded
because of problems with either the MWR data or the sounding data). The
distribution of these 1178 by facility is:
Facility Number of Sondes First Used Last Used
(based on QME)
CF 316 1/20/95 (11:30) 9/22/95 (11:30)
B1 143 2/21/95 (17:30) 9/29/95 ( 8:30)
B4 313 4/ 6/95 (17:30) 11/ 7/95 (17:28)
B5 153 4/14/95 (17:30) 12/14/95 (17:30)
B6 253 3/22/95 (17:30) 10/19/95 (17:30)
The sondes can be recognized by decoding the serial_number metadata word
in the sounding netCDF file. The serial number of the sondes in question
is a number (8 or 9 digit) of the form DDMMYTTPP where DD is the day of
the month (leading zero missing in the netCDF data), MM is the month +80
(e.g. 91 for November), Y is the last digit of the year (4 for 1994),
TT is the calibration tray ID, and PP is the position of the sonde in
the calibration tray. The sondes in question will therefore have serial
numbers coded between 26904TTPP and 22914TTPP. |
Measurements: | sgpsondewrpnC1.a1: - Relative humidity inside the instrument enclosure(rh)
sgpsondewrpnC1.a0: - Relative humidity inside the instrument enclosure(rh)
|