DQR ID | Subject | Data Streams Affected |
---|---|---|
D041027.1 | NSA/MWRP/C1 - Missing data | nsamwrpC1.b1 |
D080319.1 | NSA/MWRP/C1 - Hardware failure- Intermittent data | nsamwrpC1.b1 |
D080319.3 | NSA/MWRP/C1 - Thermal instability- TKnd test | nsamwrpC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2004 | 0925 | 02/23/2004 | 2238 |
10/07/2004 | 0859 | 10/27/2004 | 1237 |
11/20/2004 | 0039 | 11/22/2004 | 1851 |
04/10/2005 | 0000 | 04/11/2005 | 1622 |
09/04/2005 | 0531 | 09/06/2005 | 1857 |
09/10/2005 | 0052 | 09/12/2005 | 1830 |
09/21/2005 | 1805 | 10/12/2005 | 2311 |
06/23/2006 | 1800 | 08/08/2006 | 2250 |
08/11/2006 | 2359 | 08/14/2006 | 1454 |
10/09/2006 | 2101 | 10/11/2006 | 1733 |
01/05/2007 | 2358 | 01/08/2007 | 2019 |
02/28/2007 | 2309 | 03/05/2007 | 0027 |
07/13/2007 | 1838 | 12/08/2007 | 0125 |
01/04/2008 | 2357 | 01/06/2008 | 1903 |
02/16/2008 | 1818 | 02/21/2008 | 0006 |
02/23/2008 | 1005 | 02/26/2008 | 2320 |
03/12/2008 | 0143 | 04/09/2008 | 1540 |
04/10/2008 | 1840 | 04/14/2008 | 1657 |
04/18/2008 | 2258 | 04/21/2008 | 1422 |
04/27/2008 | 1846 | 04/29/2008 | 0525 |
05/01/2008 | 0328 | 05/28/2008 | 2241 |
10/14/2008 | 1659 | 02/13/2009 | 2234 |
Subject: | NSA/MWRP/C1 - Missing data |
DataStreams: | nsamwrpC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/29/2008 | 0000 | 03/12/2008 | 0000 |
Subject: | NSA/MWRP/C1 - Hardware failure- Intermittent data |
DataStreams: | nsamwrpC1.b1 |
Description: | Data are intermittent during this time due to a failure of the analog board. The calibration of V-band data may have slightly drifted due to power outages. |
Measurements: | nsamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2008 | 0000 | 03/12/2008 | 0000 |
Subject: | NSA/MWRP/C1 - Thermal instability- TKnd test |
DataStreams: | nsamwrpC1.b1 |
Description: | During this time the radiometer was thermally unstable . The noise diode physical temperature had sudden drops of more than 5 degree in several ccasions between 2/1 and 2/11. On 2/12 the radiometer temperature set point was lowered of about 10 degree as a test. On 2/27 a new analog board was installed in the instrument. Data after 2/27 show a slight change in calibration. |
Measurements: | nsamwrpC1.b1:
|