D040919.1 | NSA/GNDRAD/C1 - IRT power failure | nsagndrad60sC1.b1 |
D060209.4 | NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensors | nsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00 |
D060630.39 | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error | nsagndrad60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2004 | 0000 | 09/22/2004 | 0552 |
Subject: | NSA/GNDRAD/C1 - IRT power failure |
DataStreams: | nsagndrad60sC1.b1 |
Description: | After performing corrective maintenance which involved turning off power to the GNDRAD instruments and data logger, power to the IRT was not restored. The problem was corrected by replacing the data logger. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2006 | 0910 | 02/07/2006 | 2350 |
Subject: | NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensors |
DataStreams: | nsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00 |
Description: | During the power outage the upwelling global measurement from the PSP instrument shows a slight increase from -5 to +0 W/m2. This is due to removal of thermal offsets induced by the AC heater during the power outage. This heater is installed within the PSP ventilator - it is necessary for the NSA/C1 climate. The upwelling longwave PIR instrument is unaffected due to case and dome temperature corrections. The IRT is AC powered and was not operational during this time period. |
Measurements: | nsagndradC1.00:
nsagndrad20sC1.a0:
nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2004 | 1600 | 12/21/2005 | 1855 |
Subject: | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC1.b1:
|