DQR ID | Subject | Data Streams Affected |
---|---|---|
D030114.1 | SGP/SONDE/C1 - Bad RH Data | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D030122.1 | SGP/SONDE/C1 - Broken temperature sensor SGP/CF sounding | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D030206.5 | SGP/SONDE/C1 - Bad temperature/RH | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D060928.3 | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry | sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2003 | 0536 | 01/11/2003 | 0551 |
Subject: | SGP/SONDE/C1 - Bad RH Data |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The RH values during the first 15 minutes of this sounding oscillate between reasonable values and near-zero values. The oscillation is irregular and seems to be related to failure of the RH sensor heating process. I have informed Vaisala of the problem and provided raw data files to them for analysis. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/2003 | 1134 | 01/19/2003 | 1259 |
Subject: | SGP/SONDE/C1 - Broken temperature sensor SGP/CF sounding |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The radiosonde temperature sensor apparently broke during launch. This results in bad temperature, dew point, and rh data. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2003 | 1732 | 02/02/2003 | 1832 |
Subject: | SGP/SONDE/C1 - Bad temperature/RH |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The temperature sensor broke shortly after launch. All temperature and RH-related data are incorrect. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2006 | 1728 | 12/31/9999 | 2359 |
Subject: | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | The output precision of the following variables was increased 20060928 alt changed from 1m to 0.1m dp changed from 0.1degC to 0.01degC pres changed from 0.1 hPa to 0.01 hPa RH changed from 1%RH to 0.01%RH tdry changed from .1degC to 0.01degC Note that the increase in precision does not necessarily imply an increase in accuracy. Users are cautioned to continue to use their best scientific judgement when using these or any other data. |
Measurements: | sgpsondewnpnC1.b1:
|