DQR ID | Subject | Data Streams Affected |
---|---|---|
D060509.1 | TWP/GNDRAD/C3 - IRT failed | twpgndrad60sC3.b1 |
D060823.1 | TWP/GNDRAD/C3 - IRT bias | twpgndrad60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/2006 | 0000 | 05/18/2006 | 2359 |
Subject: | TWP/GNDRAD/C3 - IRT failed |
DataStreams: | twpgndrad60sC3.b1 |
Description: | The IRT began to fail on 4/7/2006 and failed completely (apparently due to the internal power supply) on 4/27. It was replaced on 5/18. |
Measurements: | twpgndrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2006 | 0000 | 06/27/2006 | 0344 |
Subject: | TWP/GNDRAD/C3 - IRT bias |
DataStreams: | twpgndrad60sC3.b1 |
Description: | The IRT reported higher surface temperatures than the PIR due to corrosion on the terminals of the lightning protection module and a ground wire error. The corrosion was cleaned and the wiring corrected. |
Measurements: | twpgndrad60sC3.b1:
|