DQR ID | Subject | Data Streams Affected |
---|---|---|
D080523.2 | SGP/EBBR/E2 - Net Radiation and Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
D080616.3 | SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/02/2008 | 1630 | 05/15/2008 | 2000 |
Subject: | SGP/EBBR/E2 - Net Radiation and Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The net radiometer was tipped 37 degrees eastward, resulting in apparent solar noon being 2.5 hours early. The net radiation error results in incorrect sensible and latent heat fluxes. |
Measurements: | sgp5ebbrE2.b1:
sgp15ebbrE2.b1:
sgp30ebbrE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/07/2008 | 0030 | 06/12/2008 | 1730 |
Subject: | SGP/EBBR/E2 - AEM Ventilator Dangling - Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE2.b1, sgp30ebbrE2.b1, sgp5ebbrE2.b1 |
Description: | The right AEM ventilator unit with temperature and humidity sensors inside was dangling from the AEM. Therefore, the gradients of temperature and relative humidity and thus sensible and latent heat flux are suspect for this period. Often, the fluxes look reasonable. |
Measurements: | sgp15ebbrE2.b1:
sgp5ebbrE2.b1:
sgp30ebbrE2.b1:
|