DQR ID | Subject | Data Streams Affected |
---|---|---|
D050405.1 | PYE/SKYRAD/M1 - Reprocessed: IRT configuration and calibration | pyeskyrad20sM1.a0, pyeskyrad60sM1.b1, pyeskyradM1.00 |
D050617.1 | PYE/MWR/M1 - Instrument computer locked up | pyemwrlosM1.b1 |
D050725.12 | PYE/MWR/M1 - Reprocessed: Revised Retrieval Coefficients | pyemwrlosM1.b1, pyemwrtipM1.a1 |
D050726.4 | PYE/MWR/M1 - Reprocessed: Calibration corrected | pyemwrlosM1.b1, pyemwrtipM1.a1 |
D050923.1 | PYE/SKYRAD/M1 - IRT bias | pyeskyrad60sM1.b1 |
D050927.1 | PYE/MWR/M1 - New software version (4.15) installed | pyemwrlosM1.b1, pyemwrtipM1.a1 |
D060502.2 | PYE/GNDRAD/M1 - Reprocessed: IRT configuration and calibration | pyegndrad20sM1.a0, pyegndrad60sM1.b1 |
D060630.44 | PYE/GNDRAD/M1 - Reprocessed: Longwave calibration error | pyegndrad60sM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2005 | 1821 | 03/11/2005 | 2032 |
Subject: | PYE/SKYRAD/M1 - Reprocessed: IRT configuration and calibration |
DataStreams: | pyeskyrad20sM1.a0, pyeskyrad60sM1.b1, pyeskyradM1.00 |
Description: | The configuration of the IRT was inadvertently set to default and the program of the data loggers were changed to that used at TWP.C3. This caused the calibration of the data to be wrong. It was corrected by reconfiguring the IRT and changing the conversion factors in the data logger program. These data have been reprocessed to apply correct calibrations and the reprocessed data were archived in February, 2007. |
Measurements: | pyeskyrad60sM1.b1:
pyeskyradM1.00:
pyeskyrad20sM1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2005 | 0000 | 06/09/2005 | 1500 |
Subject: | PYE/MWR/M1 - Instrument computer locked up |
DataStreams: | pyemwrlosM1.b1 |
Description: | The MWR computer locked up and required a reboot at 00Z on June 9, 2005. When the program was restarted, water vapor, liquid path, and temps dropped to zero. They did not recover until 15Z the same day. |
Measurements: | pyemwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2005 | 0700 | 05/06/2005 | 0000 |
Subject: | PYE/MWR/M1 - Reprocessed: Revised Retrieval Coefficients |
DataStreams: | pyemwrlosM1.b1, pyemwrtipM1.a1 |
Description: | IN THE BEGINNING (June 1992), the retrieval coefficients used to derive the precipitable water vapor (PWV) and liquid water path (LWP) from the MWR brightness temperatures were based on the Liebe and Layton (1987) water vapor and oxygen absorption model and the Grant (1957) liquid water absorption model. Following the SHEBA experience, revised retrievals based on the more recent Rosenkranz (1998) water vapor and oxygen absorption models and the Liebe (1991) liquid water absorption model were developed. The Rosenkranz water vapor absorption model resulted a 2 percent increase in PWV relative to the earlier Liebe and Layton model. The Liebe liquid water absorption model decreased the LWP by 10% relative to the Grant model. However, the increased oxygen absorption caused a 0.02-0.03 mm (20-30 g/m2) reduction in LWP, which was particularly significant for low LWP conditions (i.e. thin clouds encountered at SHEBA). Recently, it has been shown (Liljegren, Boukabara, Cady-Pereira, and Clough, TGARS v. 43, pp 1102-1108, 2005) that the half-width of the 22 GHz water vapor line from the HITRAN compilation, which is 5 percent smaller than the Liebe and Dillon (1969) half-width used in Rosenkranz (1998), provided a better fit to the microwave brightness temperature measurements at 5 frequencies in the range 22-30 GHz, and yielded more accurate retrievals. Accordingly, revised MWR retrieval coefficients have been developed using MONORTM, which utilizes the HITRAN compilation for its spectroscopic parameters. These new retrievals provide 3 percent less PWV and 2.6 percent greater LWP than the previous retrievals based on Rosenkranz (1998). At Point Reyes, the original coefficients implemented in March 2005 were based on a version of the Rosenkranz model that had been modified to use the HITRAN half-width at 22 GHz and to be consistent with the water vapor continuum in MONORTM. These retrievals yielded nearly identical results to the MONORTM retrievals. The MONORTM-based retrieval coefficients became active at PYE.M1 20050506. Note: The PYE.M1 MWRLOS data for 20050201-20050506 have been reprocessed to apply the MONORTM-based retrievals for all time. The reprocessed data were archived in April 2007. The TIP data have not been reprocessed. |
Measurements: | pyemwrtipM1.a1:
pyemwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/29/2005 | 0000 | 07/17/2005 | 2100 |
Subject: | PYE/MWR/M1 - Reprocessed: Calibration corrected |
DataStreams: | pyemwrlosM1.b1, pyemwrtipM1.a1 |
Description: | On May 28 1:30 GMT the NFOV radiometer was placed in the field of view of the MWR tip calibration. Almost immediately calibration of the MWR was compromised resulting in incorrect brightness temperatures and overestimation of both PWV and LWP. On July 15 the NFOV radiometer was moved away from the MWR and the instantaneous calibration values jumped back to normal. The median values returned to normal on July 17 around 2100. The LOS data were reprocessed using interpolated values for the calibration coefficients. The reprocessed data are available from the ARM Archive effective December 7, 2005. NOTE: the format of the reprocessed data are slightly different than the format of the original data and the data available before and after the reprocessed data period. The quality of the data are not affected, just the format. The MWRTIP data can not be reprocessed and should be used with caution. |
Measurements: | pyemwrtipM1.a1:
pyemwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/03/2005 | 0200 | 09/14/2005 | 2359 |
Subject: | PYE/SKYRAD/M1 - IRT bias |
DataStreams: | pyeskyrad60sM1.b1 |
Description: | The SKYRAD IRT measured lower sky temperatures than the AERI. This may have been due to a dirty mirror or lens. |
Measurements: | pyeskyrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2005 | 0700 | 09/13/2005 | 1805 |
Subject: | PYE/MWR/M1 - New software version (4.15) installed |
DataStreams: | pyemwrlosM1.b1, pyemwrtipM1.a1 |
Description: | A problem began with the installation of MWR.EXE version 4.12 in July 2002. The software had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier problem. The software upgrade corrected the earlier problem but introduced a new one that caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip curves, and saturation of CPU usage. Software versions 4.13 and 4.14 also produced these problems. The new MWR software, version 4.15, was installed on 9/13/2005. As a consequence of this upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to ~50s. |
Measurements: | pyemwrtipM1.a1:
pyemwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2005 | 1821 | 03/11/2005 | 2032 |
Subject: | PYE/GNDRAD/M1 - Reprocessed: IRT configuration and calibration |
DataStreams: | pyegndrad20sM1.a0, pyegndrad60sM1.b1 |
Description: | The configuration of the IRT was inadvertently set to default and the program of the data loggers were changed to that used at TWP.C3. This caused the calibration of the data to be wrong. It was corrected by reconfiguring the IRT and changing the conversion factors in the data logger program. These data have been reprocessed to apply correct calibrations and the reprocessed data were archived 20070212. |
Measurements: | pyegndrad60sM1.b1:
pyegndrad20sM1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2005 | 0000 | 09/14/2005 | 0000 |
Subject: | PYE/GNDRAD/M1 - Reprocessed: Longwave calibration error |
DataStreams: | pyegndrad60sM1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in February 2007. |
Measurements: | pyegndrad60sM1.b1:
|