DQR ID | Subject | Data Streams Affected |
---|---|---|
D060111.4 | SGP/EBBR/E8 - AEM Not Exchanging | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.6 | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.8 | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060501.9 | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060508.2 | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D060818.8 | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D070223.2 | SGP/EBBR/E8 - Failed Temp/RH sensor | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
D080520.1 | SGP/EBBR/E8 - Wind Direction Incorrect | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2005 | 1900 | 01/10/2006 | 1630 |
Subject: | SGP/EBBR/E8 - AEM Not Exchanging |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The AEM hung in one position frequently and sometimes for many days during this period While the AEM was not switching, the sensible and latent heat fluxes and gradient measurements were incorrect. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2005 | 0000 | 01/24/2006 | 2330 |
Subject: | SGP/EBBR/E8 - Flux and Gradient Measurements Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The flux and gradient measurements were often incorrect when the AEM did not function properly. The user needs to inspect the data (particularly the home signals) to determine when the AEM was and was not working. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2006 | 0230 | 03/21/2006 | 1600 |
Subject: | SGP/EBBR/E8 - Wind Speed Sensor Frozen Stopped |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | Freezing rain froze the wind speed sensor to a stopped condition. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2006 | 2030 | 04/04/2006 | 1500 |
Subject: | SGP/EBBR/E8 - Fluxes Incorrect; Left Humidity Near Zero |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity was intermittently incorrect, therefore the sensible and latent heat fluxes were also. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/22/2006 | 1030 | 05/31/2006 | 0030 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Flux Suspect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The left humidity sensor produced large spikes frequently. This resulted in incorrect measurements of vapor pressure gradient and thus calculations of sensible and latent heat flux. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/2006 | 0130 | 06/27/2006 | 1730 |
Subject: | SGP/EBBR/E8 - Sensible and Latent Heat Fluxes and Net Radiation Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The net radiometer top dome was broken and water had gotten into it. Sensible and latent heat fluxes and net radiation were therefore incorrect. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/22/2007 | 1630 | 03/06/2007 | 2100 |
Subject: | SGP/EBBR/E8 - Failed Temp/RH sensor |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The right Temperature-Humidity sensor failed. The measured temperature gradients were incorrect and thus the sensible and latent heat fluxes were incorrect for this extended period. The sensor was replaced March 6. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/06/2008 | 0330 | 05/28/2008 | 1745 |
Subject: | SGP/EBBR/E8 - Wind Direction Incorrect |
DataStreams: | sgp15ebbrE8.b1, sgp30ebbrE8.b1, sgp5ebbrE8.b1 |
Description: | The wind direction flatlined at times at 135 degrees (an indication of failed electronics) and at other times trended correctly with the co-located SMOS but differed from it by 15 to 100 degrees. |
Measurements: | sgp30ebbrE8.b1:
sgp5ebbrE8.b1:
sgp15ebbrE8.b1:
|