DQR ID | Subject | Data Streams Affected |
---|---|---|
D040312.1 | SGP/MMCR/C1 - Reprocess: Incorrect units for "heights" field | sgpmmcrmomC1.b1 |
D041005.2 | SGP/MMCR/C1 - Reprocessed: Corrupt CDF file | sgpmmcrmomC1.b1 |
D050117.5 | SGP/MMCR/C1 - Loss of synchronization | sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspecC1.a0 |
D050120.1 | SGP/MMCR/C1 - Instrument failure | sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspectraC1.a0 |
D050222.2 | SGP/MMCR/C1 - Power supply failure | sgparscl1clothC1.c1, sgparsclbnd1clothC1.c1, sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspectraC1.a0 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/09/2003 | 2116 | 03/01/2004 | 2359 |
Subject: | SGP/MMCR/C1 - Reprocess: Incorrect units for "heights" field |
DataStreams: | sgpmmcrmomC1.b1 |
Description: | The "heights" field metadata indicates that units are "m AGL" (meters above ground level) when in fact the heights data are recorded in "m MSL" (meters above mean sea leavel). |
Measurements: | sgpmmcrmomC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2004 | 0000 | 02/26/2004 | 2359 |
Subject: | SGP/MMCR/C1 - Reprocessed: Corrupt CDF file |
DataStreams: | sgpmmcrmomC1.b1 |
Description: | An isolated oddity in the ModeNum field of the sgpmmcrmomC1.b1 file for 2004.02.26 was identified. The following odd values appeared in the original cdf file, for 12 consecutive entries starting with profile 8611: 1047, -1, 14282, 20180, 8670, 3550, 8651, -29810, 30695, -1, 2030, 16183 The raw data were fine. This file was reprocessed on Sept 30, 2004. |
Measurements: | sgpmmcrmomC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/16/2005 | 0000 | 01/16/2005 | 1700 |
Subject: | SGP/MMCR/C1 - Loss of synchronization |
DataStreams: | sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspecC1.a0 |
Description: | During my daily review of the MMCR operaton, I noticed that the radar returns were out of synch with returns appearing all throughout the range gates. I remotely rebooted the MMCR and it returned to normal operation. |
Measurements: | sgpmmcrmomC1.b1:
sgpmmcrmomC1.00:
sgpmmcrspecC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/16/2005 | 0015 | 02/15/2005 | 2100 |
Subject: | SGP/MMCR/C1 - Instrument failure |
DataStreams: | sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspectraC1.a0 |
Description: | The receiver/modulator had a power supply that had failed. |
Measurements: | sgpmmcrmomC1.b1:
sgpmmcrmomC1.00:
sgpmmcrspectraC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/16/2005 | 0030 | 02/11/2005 | 2115 |
Subject: | SGP/MMCR/C1 - Power supply failure |
DataStreams: | sgparscl1clothC1.c1, sgparsclbnd1clothC1.c1, sgpmmcrmomC1.00, sgpmmcrmomC1.b1, sgpmmcrspectraC1.a0 |
Description: | A power supply failed in the MMCR's receiver. During the period of this DQR, the radar would appear to work but the calibration is probably incoorect during good portions of this time. The power supply was replaced and normal operation resumed at 2115 on 02/11/2005. |
Measurements: | sgparsclbnd1clothC1.c1:
sgpmmcrmomC1.b1:
sgpmmcrmomC1.00:
sgpmmcrspectraC1.a0:
sgparscl1clothC1.c1:
|