DQR ID | Subject | Data Streams Affected |
---|---|---|
D040831.4 | SGP/SIRS/E10 - Reprocess: Incorrect shortwave radiometer calibration coefficients | sgpsirsE10.00, sgpsirsE10.b1 |
D041206.1 | SGP/SIRS/E10 - tracker failure | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
D060309.1 | SGP/SIRS/E1 - Reprocess: incorrect calibration coefficients for US (upwelling shortwave) instrument | sgpsirsE1.00, sgpsirsE1.b1 |
D060331.2 | SGP/SIRS/E1 - Intermittantly missing downwelling diffuse | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
D060630.18 | SGP/SIRS/E10 - Reprocessed: Longwave Calibration error | sgpsirsE10.b1 |
D060630.9 | SGP/SIRS/E1 - Reprocessed: Longwave Calibration error | sgpsirsE1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 1630 | 04/21/2004 | 1425 |
Subject: | SGP/SIRS/E10 - Reprocess: Incorrect shortwave radiometer calibration coefficients |
DataStreams: | sgpsirsE10.00, sgpsirsE10.b1 |
Description: | During the pyrgeometer instrument (PIR) changeout a datalogger program was loaded with incorrect calibration coefficients for the shortwave instruments. The affected measurements (instruments) were: downwelling diffuse (8-48), upwelling shortwave (PSP), direct normal (NIP) and downwelling shortwave (PSP) Recalculate irradiances by determining original mV signal and reapplying the correct calibration coefficients. For SGP/SIRS/E10 and each one minute irradiance value perform the following calculations by instrument: DD(corrected) = (DD/107.24)*114.63 US(corrected) = (US/113.19)*117.46 NIP(corrected)= (NIP/120.17)*119.26 DS(corrected) = (DS/118.46)*122.78 |
Measurements: | sgpsirsE10.b1:
sgpsirsE10.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2004 | 1800 | 01/26/2005 | 1600 |
Subject: | SGP/SIRS/E10 - tracker failure |
DataStreams: | sgpsirs20sE10.a0, sgpsirsE10.00, sgpsirsE10.b1 |
Description: | During the night of 11/05-06 2004 the tracker failed at this location. During CM on 12/01, power cycling returned the tracker to service. However, tracker was not properly tracking in the AM. Tracker failed again sometime near 01/01/2005. CM on reloaded the tracker program and realigned the tracker on 01/26/2005 1600GMT. It has been working correctly since this time. |
Measurements: | sgpsirsE10.b1:
sgpsirsE10.00:
sgpsirs20sE10.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2006 | 2008 | 03/07/2006 | 1950 |
Subject: | SGP/SIRS/E1 - Reprocess: incorrect calibration coefficients for US (upwelling shortwave) |
DataStreams: | sgpsirsE1.00, sgpsirsE1.b1 |
Description: | During the changeover from 4 to 1 coefficient for the PIRs at SIRS/E1 (PIF) the datalogger program was programmed with the wrong calibration coefficient and serial number for the US PSP instrument measuring the Upwelling Hemispheric Shortwave radiation. The calibration coefficient was 114.43 W/m2/mV (instrument 30618F3) when it should have been 104.47W/m2/mV (instrument 29618F3) |
Measurements: | sgpsirsE1.b1:
sgpsirsE1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/26/2005 | 1630 | 04/04/2006 | 2005 |
Subject: | SGP/SIRS/E1 - Intermittantly missing downwelling diffuse |
DataStreams: | sgpsirs20sE1.a0, sgpsirsE1.00, sgpsirsE1.b1 |
Description: | Starting on or about 10/26/2005 the downwelling diffuse hemispheric irradiance (Eppley 8-48) instrument had intermittant missing values. Technicians found a loose wire on the instrument connector and repaired it on 4/4/2006. Use global shortwave hemispheric (PSP) component - normal incident component (CosZA*NIP) to calulate diffuse component. Comparison of absolute difference between actual and computed diffuse component values and diffuse standard deviations > 1-2 W/m2 can help filter the noisy diffuse data. There is no standard method. |
Measurements: | sgpsirs20sE1.a0:
sgpsirsE1.b1:
sgpsirsE1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 1730 | 02/07/2006 | 2130 |
Subject: | SGP/SIRS/E10 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE10.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE10.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2004 | 2055 | 02/21/2006 | 2007 |
Subject: | SGP/SIRS/E1 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE1.b1:
|