DQR ID | Subject | Data Streams Affected |
---|---|---|
D040428.1 | TWP/SMET/C1 - Poor Performing Sensor | twpsmet60sC1.b1 |
D050210.1 | TWP/SMET/C3 - Data non-representative | twpsmet60sC3.b1 |
D050301.1 | TWP/SMET/C2 - Instrument failure | twpsmet60sC2.b1 |
D050322.2 | TWP/SMET/C2 - Upper wind sensor failure | twpsmet60sC2.b1 |
D050418.2 | TWP/SMET/C3 - Instrument logger problem, Sensor failure | twpsmet60sC3.b1 |
D050714.1 | TWP/SMET/C2 - RH failure | twpsmet60sC2.b1 |
D050808.3 | TWP/SMET/C3 - Data not representative | twpsmet60sC3.b1 |
D050831.4 | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative | twpsmet60sC3.b1 |
D050920.4 | TWP/SMET/C2 - Instrument maintenance | twpsmet60sC2.b1 |
D050921.2 | TWP/SMET/C2 - ORG data incorrect | twpsmet60sC2.b1 |
D051005.6 | TWP/SMET/C2 - Tower down for calibration activities | twpsmet60sC2.b1 |
D051115.1 | TWP/SMET/C3 - Upper wind sensor failure | twpsmet60sC3.b1 |
D051117.2 | TWP/SMET/C3 - RH Values Artificially High | twpsmet60sC3.b1 |
D051117.3 | TWP/SMET/C3 - Instrument failure | twpsmet60sC3.b1 |
D060426.8 | TWP/SMET/C3 - ORG data noisy due to degraded ground | twpsmet60sC3.b1 |
D060502.3 | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error | twpsmet60sC3.b1 |
D060522.1 | TWP/SMET/C1 - Upper wind speed sensor failure | twpsmet60sC1.b1 |
D060614.2 | TWP/SMET/C2 - Temp/RH probe failure | twpsmet60sC2.b1 |
D060627.1 | TWP/SMET/C3 - Anemometer Readings Incorrect | twpsmet60sC3.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
D060818.2 | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC2.b1 |
D060818.3 | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC3.b1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
D061011.3 | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC2.b1 |
D061011.4 | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC3.b1 |
D070306.2 | TWP/SMET/C3 - Barometer Data Incorrect | twpsmet60sC3.b1 |
D070306.3 | TWP/SMET/C3 - Data Incorrect Due to Water Intrusion | twpsmet60sC3.b1 |
D070409.1 | TWP/SMET/C1 - Upper winds failure | twpsmet60sC1.b1 |
D070510.5 | TWP/SMET/C2 - Lower windspeed sensor questionable | twpsmet60sC2.b1 |
D070520.1 | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect | twpsmet60sC2.b1 |
D070803.1 | TWP/SMET/C3 - Upper wind speed artificially slow | twpsmet60sC3.b1 |
D071105.1 | TWP/SMET/C3 - ORG overestimated rainfall | twpsmet60sC3.b1 |
D071219.1 | TWP/SMET/C2 - Lower Wind Speed Data Suspect | twpsmet60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 0000 | 01/19/2005 | 1530 |
Subject: | TWP/SMET/C1 - Poor Performing Sensor |
DataStreams: | twpsmet60sC1.b1 |
Description: | On 4/23/2004 condensation was found inside the target area of the ORG lens. A spare was sent and installed but did not solve the problem of high background voltage values causing rainrate values to be recorded during no rainfall. Values range anywhere from 0.06 to 0.12 mm/hr. A new Optical Rain Gage was installed on 01/19/2005 @ 13:30 GMT alleviating the high background voltage values that lead to consistent rainfall rates of around 0.11 to 0.15 mm/hr on a consistent basis. The formula used to calculate rainfall rates from the new Optical Rain Gage is: RR (mm/hr) = 25(V^1.87) - 0.15. Older model formula = 20(V^2) - 0.05 The two formulas result in comparable data. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2005 | 2310 | 02/09/2005 | 2320 |
Subject: | TWP/SMET/C3 - Data non-representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower mast was lowered for inspection. Data during this time is non-representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2005 | 1955 | 03/03/2005 | 0430 |
Subject: | TWP/SMET/C2 - Instrument failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Upper wind sensor failed. Wind direction was reported at a near constant value of 69 degrees and was significantly different than the lower wind sensor's direction values. Sensor replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2005 | 2300 | 04/17/2005 | 0332 |
Subject: | TWP/SMET/C2 - Upper wind sensor failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | After replacement on 03/03/2005 the upper wind sensor began reporting erroneous values again on 03/09/2005 becoming very apparent after 03/17/2005. Wind speed data appears reasonable but the wind direction data do not agree well with the lower wind sensor. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2005 | 0230 | 04/19/2005 | 0050 |
Subject: | TWP/SMET/C3 - Instrument logger problem, Sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | The wind_lower direction sensor failed on 4/14 and disagrees significantly with the wind_upper direction. The wind_lower sensor reports a constant value around 0-15 degrees since it failed. Testing showed that excitation channel two on the logger was the problem. Logger was replaced by spare. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2005 | 1900 | 07/14/2005 | 2226 |
Subject: | TWP/SMET/C2 - RH failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | The RH values were frequently over MAX value of 104% starting on 7/10/2005 through end of period. Sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower was lowered for radiometer change and characterization. Wind, temp and RH data not representative of local conditions. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | The tower was lowered for radiometer change and comparisons. During the time period the data listed was not representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/2005 | 1625 | 09/21/2005 | 1815 |
Subject: | TWP/SMET/C2 - Instrument maintenance |
DataStreams: | twpsmet60sC2.b1 |
Description: | Instrument calibration and testing were taking place for brief periods. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2005 | 0809 | 09/19/2005 | 1902 |
Subject: | TWP/SMET/C2 - ORG data incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Data from the ORG was incorrect due to heavy condensation on the lens. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2005 | 0200 | 09/19/2005 | 0330 |
Subject: | TWP/SMET/C2 - Tower down for calibration activities |
DataStreams: | twpsmet60sC2.b1 |
Description: | Tower lowered, wind vanes spun at different frequencies, and ORG tested with electronic rain. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 2215 |
Subject: | TWP/SMET/C3 - Upper wind sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike destroyed the upper (#1) wind sensor. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 0050 |
Subject: | TWP/SMET/C3 - RH Values Artificially High |
DataStreams: | twpsmet60sC3.b1 |
Description: | Lightning strike caused RH readings to be higher than reasonable (130%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/23/2005 | 0000 |
Subject: | TWP/SMET/C3 - Instrument failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike caused a failure in the ORG. Rainfall was being continually recorded. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2005 | 0000 | 08/24/2006 | 2359 |
Subject: | TWP/SMET/C3 - ORG data noisy due to degraded ground |
DataStreams: | twpsmet60sC3.b1 |
Description: | SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse over time so that in February 2006 the ORG was reporting with a constant background value of around 0.10 mm/hr rain rate. ORG voltage showed a near constant value above the 85mV threshold. The ground for the ORG had become degraded and was lost due to corrosion. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/27/2006 | 2330 | 05/02/2006 | 0201 |
Subject: | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error |
DataStreams: | twpsmet60sC3.b1 |
Description: | The lower anemometer was replaced afer being removed in anticipation of cyclone Monica. The calibration coefficient was not changed during this period so the wind speeds are a little low (0.2%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2006 | 0000 | 10/14/2006 | 0237 |
Subject: | TWP/SMET/C1 - Upper wind speed sensor failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind speed sensor was binding at low wind speeds. Comparison to the lower wind sensor showed that the upper sensor flat-lines for long periods of time. The upper anemometer was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2006 | 2144 | 06/15/2006 | 0427 |
Subject: | TWP/SMET/C2 - Temp/RH probe failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Temperature/RH probe failed and was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2006 | 0115 | 06/27/2006 | 0344 |
Subject: | TWP/SMET/C3 - Anemometer Readings Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The Anemometer readings were incorrect due to the tower being lowered for maintenance. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 07/13/2006 | 2232 |
Subject: | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC2.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2003 | 0000 | 07/13/2006 | 2219 |
Subject: | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC3.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 10/17/2006 | 0000 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC1.b1 datastream. Between 9/26 and 10/17, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19961009-20060926 and data from 19961009-20061017 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 11/29/2006 | 2014 |
Subject: | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC2.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C2 SMET suite of instruments on 20061129. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC2.b1 datastream. Between 9/26 and 11/29, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19981021-20060926 and data from 19981021-20061129 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0000 | 09/26/2006 | 1435 |
Subject: | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC3.b1 |
Description: | A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 datastream. When reprocessing of this datastream occurs these variables will be added with -9999 (missing) as the value for dates prior to gauge installation. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2007 | 1723 | 03/03/2007 | 1228 |
Subject: | TWP/SMET/C3 - Barometer Data Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The barometer data was incorrect or missing due to water intrusion. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2007 | 0056 | 03/03/2007 | 0123 |
Subject: | TWP/SMET/C3 - Data Incorrect Due to Water Intrusion |
DataStreams: | twpsmet60sC3.b1 |
Description: | Data became erratic and unreasonable after water intrusion into the enclosure. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2007 | 0000 | 06/12/2007 | 0406 |
Subject: | TWP/SMET/C1 - Upper winds failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind instrument was inoperative resulting in constant wind direction and speed. A failed cable was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2007 | 0000 | 08/19/2007 | 2135 |
Subject: | TWP/SMET/C2 - Lower windspeed sensor questionable |
DataStreams: | twpsmet60sC2.b1 |
Description: | Lower wind speed sensor had degraded bearings. This sensor was reporting as much as 0.4 m/s lower values than the upper sensors when the winds were 2 m/s or below. The sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/11/2007 | 0000 | 05/22/2007 | 0047 |
Subject: | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Wind directions between lower and upper anemometers were displaced by approximately 45 degrees from 05/10/2007 through 05/21/2007 when a new sensor was installed to replace a failing sensor. The new sensor direction data incorrect. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2007 | 0000 | 09/18/2007 | 2315 |
Subject: | TWP/SMET/C3 - Upper wind speed artificially slow |
DataStreams: | twpsmet60sC3.b1 |
Description: | The upper wind measurement was consistently slower by around 0.5 - 1 m/s than the lower wind measurement. The bearings on the upper wind measurement were binding. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/2007 | 1855 | 11/02/2007 | 2010 |
Subject: | TWP/SMET/C3 - ORG overestimated rainfall |
DataStreams: | twpsmet60sC3.b1 |
Description: | Optical rain gauge (ORG) measured approximately 10mm higher than two co-located tipping bucket rain gauges and a co-located disdrometer. The ORG is typically higher but usually not by an order of magnitude. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/2007 | 0048 | 08/19/2007 | 2150 |
Subject: | TWP/SMET/C2 - Lower Wind Speed Data Suspect |
DataStreams: | twpsmet60sC2.b1 |
Description: | The lower wind speed bearings appeared to be binding at speeds below 2 m/s as it did not compare well with the upper wind monitor. |
Measurements: | twpsmet60sC2.b1:
|