DQR ID | Subject | Data Streams Affected |
---|---|---|
D031216.1 | NSA/GNDRAD/C1 - IRT cable failure | nsagndrad60sC1.b1 |
D031216.2 | NSA/GNDRAD/C1 - IRT returned to vendor | nsagndrad60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/2003 | 1400 | 09/11/2003 | 1906 |
Subject: | NSA/GNDRAD/C1 - IRT cable failure |
DataStreams: | nsagndrad60sC1.b1 |
Description: | The IRT signal became very noisy and the problem was traced to the cable (after the instrument was returned to the vendor for diagnosis). The problem started occurring when ice surrounding the cable began to melt. The cable was repaired and the instrument reinstalled. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2003 | 1906 | 10/15/2003 | 0015 |
Subject: | NSA/GNDRAD/C1 - IRT returned to vendor |
DataStreams: | nsagndrad60sC1.b1 |
Description: | The IRT was returned to the vendor for diagnosis. It was found to be operating as expected and the signal noise problem was traced to the cable. The cable was repaired and the instrument reinstalled. |
Measurements: | nsagndrad60sC1.b1:
|