DQR ID | Subject | Data Streams Affected |
---|---|---|
D030730.1 | TWP/GNDRAD/C2 - IRT problem | twpgndrad60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/12/2003 | 2100 | 09/14/2003 | 1100 |
Subject: | TWP/GNDRAD/C2 - IRT problem |
DataStreams: | twpgndrad60sC2.b1 |
Description: | The GNDRAD IRT exhibited a positive bias compared to the PIR from 2100 to 0700 UTC each day. The problem was corrected by removing the signal cable, cleaning the contacts, and reconnecting the cable. |
Measurements: | twpgndrad60sC2.b1:
|