DQR ID | Subject | Data Streams Affected |
---|---|---|
D070628.1 | FKB/MWRP/M1 - Reprocess: Incorrect calibration | fkbmwrpM1.b1 |
D071009.1 | FKB/MWRP/M1 - LN2 Calibration- | fkbmwrpM1.b1 |
D071016.1 | FKB/MWRP/M1 - Reprocess: Updated 30-GHz calibration coefficient | fkbmwrpM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/20/2007 | 0520 | 06/21/2007 | 2000 |
Subject: | FKB/MWRP/M1 - Reprocess: Incorrect calibration |
DataStreams: | fkbmwrpM1.b1 |
Description: | The software was started with the wrong calibration data. |
Measurements: | fkbmwrpM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2007 | 1100 | 09/17/2007 | 1500 |
Subject: | FKB/MWRP/M1 - LN2 Calibration- |
DataStreams: | fkbmwrpM1.b1 |
Description: | Between 1100 and 1500 there were two calibration attempts. |
Measurements: | fkbmwrpM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2007 | 1500 | 10/09/2007 | 1840 |
Subject: | FKB/MWRP/M1 - Reprocess: Updated 30-GHz calibration coefficient |
DataStreams: | fkbmwrpM1.b1 |
Description: | On 9/17 the instrument was calibrated with liquid nitrogen. For the 30-GHz channel the results of calibration were not optimal and were causing a negative bias in the retrieved LWP. On 10/9 the 30-GHz calibration parameters were updated with values derived from tip curves. Data between 9/17 and 10/9 will be reprocessed to correct for the 30-GHz calibration. |
Measurements: | fkbmwrpM1.b1:
|