DQR ID | Subject | Data Streams Affected |
---|---|---|
D040812.2 | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
D051202.5 | TWP/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 downwelling longwave | twpskyrad20sC1.a0, twpskyrad60sC1.b1 |
D051202.6 | TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 downwelling longwave | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
D060630.31 | TWP/SKYRAD/C1 - Reprocess: Longwave Calibration error | twpskyrad60sC1.b1 |
D060630.32 | TWP/SKYRAD/C2 - Reprocess: Longwave Calibration error | twpskyrad60sC2.b1 |
D060630.34 | TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error | twpgndrad60sC1.b1 |
D060630.35 | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error | twpgndrad60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1996 | 1934 | 08/01/2004 | 2359 |
Subject: | TWP/SKYRAD/C1 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC1.a1, twpskyrad60sC1.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC1.b1:
twpskyrad20sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2004 | 0200 | 09/09/2005 | 2022 |
Subject: | TWP/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 |
DataStreams: | twpskyrad20sC1.a0, twpskyrad60sC1.b1 |
Description: | The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were installed. The result of no delay in these measurements results in an error in the measured resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 irradiances. The absolute value of the errors are inversely proportial to temperature. Users can apply temperature dependent irradiance corrections shown in the table below. Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 W/m2. The error corrections become the same order of magnitude or larger than the instrument uncertainties at or below 260K or 9F. Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance DIR1 DIR1 DIR2 DIR2 T[F] Tcase[K] Correction[W/m2] Tdome[K] Correction[W/m2] 109.4 316.13 0.00 316.13 0.00 100.5 311.20 -0.15 311.20 0.12 93.2 307.14 -0.12 307.14 0.10 87.0 303.70 -0.10 303.70 0.08 81.6 300.70 -0.17 300.70 0.14 77.0 298.15 -0.22 298.15 0.18 59.4 288.35 -0.34 288.35 0.27 47.4 281.72 -0.23 281.72 0.18 38.5 276.74 -0.42 276.74 0.34 31.3 272.77 -0.48 272.77 0.38 25.4 269.48 -0.61 269.48 0.49 20.3 266.67 -0.67 266.67 0.54 16.0 264.25 -0.82 264.25 0.66 12.1 262.11 -0.98 262.11 0.79 8.7 260.19 -0.97 260.19 0.78 5.6 258.46 -1.07 258.46 0.86 2.7 256.89 -1.17 256.89 0.94 0.1 255.44 -1.33 255.44 1.07 -2.3 254.11 -1.41 254.11 1.13 -4.5 252.87 -1.43 252.87 1.15 -6.6 251.71 -1.52 251.71 1.22 -8.5 250.63 -1.63 250.63 1.31 -10.4 249.62 -1.73 249.62 1.39 -12.2 248.58 -1.79 248.58 1.44 -15.4 246.82 -1.93 246.82 1.55 -18.3 245.23 -2.05 245.23 1.65 -20.9 243.78 -2.22 243.78 1.78 -23.3 242.44 -2.36 242.44 1.89 -25.5 241.21 -2.58 241.21 2.07 -27.5 240.07 -2.65 240.07 2.13 -29.5 239.01 -2.75 239.01 2.21 -31.3 238.00 -3.02 238.00 2.42 |
Measurements: | twpskyrad20sC1.a0:
twpskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 2200 | 09/09/2005 | 2058 |
Subject: | TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were installed. The result of no delay in these measurements results in an error in the measured resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 irradiances. The absolute value of the errors are inversely proportial to temperature. Users can apply the temperature dependent irradiance corrections shown in the table below. Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 W/m2. The error corrections become the same order of magnitude or larger than the instrument uncertainties at or below 260K or 9F. Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance DIR1 DIR1 DIR2 DIR2 T[F] Tcase[K] Correction[W/m2] Tdome[K] Correction[W/m2] 109.4 316.13 0.00 316.13 0.00 100.5 311.20 -0.15 311.20 0.12 93.2 307.14 -0.12 307.14 0.10 87.0 303.70 -0.10 303.70 0.08 81.6 300.70 -0.17 300.70 0.14 77.0 298.15 -0.22 298.15 0.18 59.4 288.35 -0.34 288.35 0.27 47.4 281.72 -0.23 281.72 0.18 38.5 276.74 -0.42 276.74 0.34 31.3 272.77 -0.48 272.77 0.38 25.4 269.48 -0.61 269.48 0.49 20.3 266.67 -0.67 266.67 0.54 16.0 264.25 -0.82 264.25 0.66 12.1 262.11 -0.98 262.11 0.79 8.7 260.19 -0.97 260.19 0.78 5.6 258.46 -1.07 258.46 0.86 2.7 256.89 -1.17 256.89 0.94 0.1 255.44 -1.33 255.44 1.07 -2.3 254.11 -1.41 254.11 1.13 -4.5 252.87 -1.43 252.87 1.15 -6.6 251.71 -1.52 251.71 1.22 -8.5 250.63 -1.63 250.63 1.31 -10.4 249.62 -1.73 249.62 1.39 -12.2 248.58 -1.79 248.58 1.44 -15.4 246.82 -1.93 246.82 1.55 -18.3 245.23 -2.05 245.23 1.65 -20.9 243.78 -2.22 243.78 1.78 -23.3 242.44 -2.36 242.44 1.89 -25.5 241.21 -2.58 241.21 2.07 -27.5 240.07 -2.65 240.07 2.13 -29.5 239.01 -2.75 239.01 2.21 -31.3 238.00 -3.02 238.00 2.42 |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2004 | 0200 | 12/21/2005 | 2205 |
Subject: | TWP/SKYRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | twpskyrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 0500 | 12/21/2005 | 2150 |
Subject: | TWP/SKYRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | twpskyrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2004 | 0700 | 12/21/2005 | 2204 |
Subject: | TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 2202 | 12/21/2005 | 2149 |
Subject: | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC2.b1:
|