DQR ID | Subject | Data Streams Affected |
---|---|---|
D061117.1 | NIM/MWRP/M1 - Reprocess: K-band calibration drift | nimmwrpM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2006 | 0000 | 11/17/2006 | 1713 |
Subject: | NIM/MWRP/M1 - Reprocess: K-band calibration drift |
DataStreams: | nimmwrpM1.b1 |
Description: | During October and November there was a drift in the K-Band calibration of the MWRP. This resulted in brightness temperatures that are too high. New calibration coefficients were derived from the median value of the noise diode temperature for the months of October and November. |
Measurements: | nimmwrpM1.b1:
|