DQR ID | Subject | Data Streams Affected |
---|---|---|
D040428.1 | TWP/SMET/C1 - Poor Performing Sensor | twpsmet60sC1.b1 |
D040824.3 | TWP/SMET/C1 - Incorrect rainrate equation used | twpsmet60sC1.b1 |
D050421.6 | TWP/SMET/C1 - Negative rainfall rates | twpsmet60sC1.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 0000 | 01/19/2005 | 1530 |
Subject: | TWP/SMET/C1 - Poor Performing Sensor |
DataStreams: | twpsmet60sC1.b1 |
Description: | On 4/23/2004 condensation was found inside the target area of the ORG lens. A spare was sent and installed but did not solve the problem of high background voltage values causing rainrate values to be recorded during no rainfall. Values range anywhere from 0.06 to 0.12 mm/hr. A new Optical Rain Gage was installed on 01/19/2005 @ 13:30 GMT alleviating the high background voltage values that lead to consistent rainfall rates of around 0.11 to 0.15 mm/hr on a consistent basis. The formula used to calculate rainfall rates from the new Optical Rain Gage is: RR (mm/hr) = 25(V^1.87) - 0.15. Older model formula = 20(V^2) - 0.05 The two formulas result in comparable data. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2004 | 0307 | 08/17/2004 | 2138 |
Subject: | TWP/SMET/C1 - Incorrect rainrate equation used |
DataStreams: | twpsmet60sC1.b1 |
Description: | There are two different Optical Raingages in use at the TWP sites. The incorrect equation was used at this site. The incorrect equation is (25*(V^1.87)) - 0.15. The correct equation is (20*(V^2)) - 0.05. The correct equation was installed. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 03/09/2004 | 2243 |
Subject: | TWP/SMET/C1 - Negative rainfall rates |
DataStreams: | twpsmet60sC1.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. New loggers were installed 03/09/2004 @ 2243 GMT. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 09/26/2006 | 1438 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 datastream. When reprocessing of this datastream occurs these variables will be added with -9999 (missing) as the value for dates prior to gauge installation. |
Measurements: | twpsmet60sC1.b1:
|