DQR ID | Subject | Data Streams Affected |
---|---|---|
D040428.1 | TWP/SMET/C1 - Poor Performing Sensor | twpsmet60sC1.b1 |
D040824.3 | TWP/SMET/C1 - Incorrect rainrate equation used | twpsmet60sC1.b1 |
D050601.4 | TWP/SMET/C1 - Reprocess: Incorrect Units Listed for Vapor Pressure | twpsmet60sC1.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 0000 | 01/19/2005 | 1530 |
Subject: | TWP/SMET/C1 - Poor Performing Sensor |
DataStreams: | twpsmet60sC1.b1 |
Description: | On 4/23/2004 condensation was found inside the target area of the ORG lens. A spare was sent and installed but did not solve the problem of high background voltage values causing rainrate values to be recorded during no rainfall. Values range anywhere from 0.06 to 0.12 mm/hr. A new Optical Rain Gage was installed on 01/19/2005 @ 13:30 GMT alleviating the high background voltage values that lead to consistent rainfall rates of around 0.11 to 0.15 mm/hr on a consistent basis. The formula used to calculate rainfall rates from the new Optical Rain Gage is: RR (mm/hr) = 25(V^1.87) - 0.15. Older model formula = 20(V^2) - 0.05 The two formulas result in comparable data. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2004 | 0307 | 08/17/2004 | 2138 |
Subject: | TWP/SMET/C1 - Incorrect rainrate equation used |
DataStreams: | twpsmet60sC1.b1 |
Description: | There are two different Optical Raingages in use at the TWP sites. The incorrect equation was used at this site. The incorrect equation is (25*(V^1.87)) - 0.15. The correct equation is (20*(V^2)) - 0.05. The correct equation was installed. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2004 | 0000 | 05/19/2005 | 2359 |
Subject: | TWP/SMET/C1 - Reprocess: Incorrect Units Listed for Vapor Pressure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The netcdf file header incorrectly listed the units of Vapor Pressure as hPa when in fact they were kPa. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|