Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/2010 | 2159 | 06/17/2010 | 2201 |
Subject: | NSA/VCEIL/C1 - Gaps in data |
DataStreams: | nsavceil25kC1.b1 |
Description: | When the CT25K ceilometer was replaced with a new CL31, checksum errors began occurring at regular 6-min intervals and several things were done to diagnose the problem, including changing the serial cable and trying different computer types. It was finally corrected by replacing the computer with one in which hyperthreading was disabled by default in the BIOS. |
Suggestions: | |
Measurements: | nsavceil25kC1.b1: more
|