Subject: | GRW/MWRP/M1 - Failed LN2 calibration |
DataStreams: | grwmwrpM1.b1
|
Description: | On 4/28 we tried to perform a LN2 calibration. Because of high humidity and lack of heater
the calibration was unsuccessful and wrong calibration coefficients were uploaded in the
configuration file. At 14:10 UTC the old configuration file was reinstated. |
Suggestions: | Data between 8:10 and 14:10 UTC should be disregarded |
Measurements: | grwmwrpM1.b1: - Derived relative humidity(relativeHumidity)
- Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
- Retrieved cloud liquid water content(liquidWaterContent)
- Interpolated water vapor mixing ratio(waterVaporMixingRatio)
- Microwave brightness temperature(brightnessTemperature)
- Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
- Retrieved liquid water path(liquidWaterPath)
- Retrieved water vapor density(waterVaporDensity)
- Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
- Derived virtual temperature(virtualTemperature)
- Temperature used for calculation(temperature)
more
|