Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2007 | 0000 | 06/26/2007 | 2025 |
Subject: | SGP/SMOS/E1 - Failed temperature probe |
DataStreams: | sgp30smosE1.b1, sgp1440smosE1.b1, sgp1smosE1.b1 |
Description: | The temperature probe was out of tolerance as it failed the 6-month calibration checks and subsequent bi-weekly checks. The temperature values consistently were outside the manufacturers suggested uncertainty of 0.2 C. |
Suggestions: | Use temperature data from other nearby SMOS sites. |
Measurements: | sgp30smosE1.b1: more
sgp1440smosE1.b1: more sgp1smosE1.b1: more |