Subject: | SGP/SURTHREF/C1 - Instrument failure on Vaisala probe 1 and probe 3. |
DataStreams: | sgpsurthrefC1.b1
|
Description: | For the dates: 11/01/2005 - 12/01/2005 T/RH probe Vaisala #3 failed. RH data was missing
and the temp data wass higher than other probes. Corrosion on the connectors caused the
problem.
For the dates: 12/03/05 - 12/12/2005 T/RH probe Vaisala #1 Temperatures were lower than
other sensors and intermittently failed. A loose wire was found and corrected. |
Suggestions: | For both instances, remove these probes data from any usage during these time periods. |