Subject: | SGP/SIRS/E22 - Intermittent upwelling longwave data loss |
DataStreams: | sgpsirs20sE22.a0, sgpsirsE22.00, sgpsirsE22.b1
|
Description: | Upwelling longwave hemispheric data drops out occasionally during the afternoon hours.
Maintenance determined that the Upwelling LW Hemispheric irradiance instrument, the PIR,
had a failed cable affecting only the thermopile measurement of this instrument. Cable was
replaced 7/20/05 2000GMT. |
Suggestions: | Intermittent missing data during this time range |
Measurements: | sgpsirsE22.00: - (Raw data stream - documentation not supported)
more
sgpsirsE22.b1: - Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
Standard Deviation(up_long_hemisp_std) - Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_netir)
- Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
- Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
- 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
more
sgpsirs20sE22.a0: - Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
- Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance Pyrgeometer(inst_up_long_dome_resist)
- Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
more
|