Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2005 | 1821 | 03/11/2005 | 2032 |
Subject: | PYE/SKYRAD/M1 - Reprocessed: IRT configuration and calibration |
DataStreams: | pyeskyradM1.00, pyeskyrad60sM1.b1, pyeskyrad20sM1.a0 |
Description: | The configuration of the IRT was inadvertently set to default and the program of the data loggers were changed to that used at TWP.C3. This caused the calibration of the data to be wrong. It was corrected by reconfiguring the IRT and changing the conversion factors in the data logger program. These data have been reprocessed to apply correct calibrations and the reprocessed data were archived in February, 2007. |
Suggestions: | Retrieve reprocessed data from ARM Archive. |
Measurements: | pyeskyrad60sM1.b1: more
pyeskyradM1.00: more pyeskyrad20sM1.a0: more |