Subject: | NSA/GNDRAD/C1 - IRT data noisy |
DataStreams: | nsagndrad60sC1.b1, nsagndrad60sC1.a1
|
Description: | After re-installation of the IRT on 12/10/99, the data are noisy. Source of the noise is
believed to be the heater connected directly to the IRT. Instrument was taken offline on
12/30/1999 and shipped back to PNNL for repair/testing. |
Suggestions: | Use caution when using this data. |
Measurements: | nsagndrad60sC1.b1: - Surface infra-red temperature(sfc_ir_temp)
- Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
- Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)
- Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
more
nsagndrad60sC1.a1: - IRT1 mean(irt1_mean)
- IRT1 maximum(irt1_max)
- IRT1 standard deviation(irt1_sd)
- IRT1 minimum(irt1_min)
more
|