Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/01/2005 | 0000 | 03/28/2006 | 2359 | Does not affect quality |
Subject: | SGP/IRT/E11 - Response time |
DataStreams: | sgpirt200msE11.a1, sgpirtE11.b1 |
Description: | The time response of the temperature measurement was changed from 1.0 to 0.1 sec to accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a sky temperature of -50C) and a significant decrease in the signal-to-noise ratio. |
Suggestions: | |
Measurements: | sgpirtE11.b1: more
sgpirt200msE11.a1: more |