Data Quality Reports for Session: 182631 User: gnk Completed: 01/31/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160621.4ASI/AOS/M1 - Manifold blower not workingasiaossp2auxM1.00, asiaossp2M1.00Incorrect
D161031.5ASI/SP2/M1 - Transferasiaossp2M1.00, asiaossp2auxM1.00Missing


DQRID : D160621.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/16/2016004206/16/20161345Incorrect
Subject:
ASI/AOS/M1 - Manifold blower not working
DataStreams:asiaossp2auxM1.00, asiaossp2M1.00
Description:
Manifold blower not working, instrument sample flows either zero or different than 
specified. Sampling through the stack not isokinetic.
Measurements:asiaossp2M1.00:
  • Raw data stream - documentation not supported

asiaossp2auxM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D161031.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/26/2016150010/27/20161711Missing
Subject:
ASI/SP2/M1 - Transfer
DataStreams:asiaossp2M1.00, asiaossp2auxM1.00
Description:
The SP2 was turned off for the specified time period as a precaution brought about by 
problems with the two generators. Once the site power was stable, the SP2 was turned back on
Measurements:asiaossp2M1.00:
  • Raw data stream - documentation not supported

asiaossp2auxM1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA