Data Quality Reports for Session: 182629 User: koontz Completed: 01/06/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160923.2OLI/PSAP/M1 - Data gaps from pump turned offoliaospsaprM1.a0Incorrect
D161211.1OLI/PSAP/M1 - Incorrect reset of transmittance after filter changeoliaospsaprM1.a0Incorrect
D161211.2OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, power failuresoliaospsaprM1.a0Incorrect
D161211.3OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, etc. oliaospsaprM1.a0Incorrect
D161211.4OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned offoliaospsaprM1.a0Incorrect


DQRID : D160923.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/06/2016155108/06/20161808Incorrect
08/18/2016225208/20/20162302Incorrect
08/25/2016192508/31/20162359Incorrect
08/21/2016231108/22/20161351Incorrect
Subject:
OLI/PSAP/M1 - Data gaps from pump turned off
DataStreams:oliaospsaprM1.a0
Description:
Data Gaps > 1h due to Tr<0.7, pump turned off, etc. 
New Operator replaced previous Operator and new Operator unaware of need to change the 
filter.
Measurements:oliaospsaprM1.a0:
  • red_overflow_count
  • blue_signal_sum
  • green_overflow_count
  • dark_overflow_count
  • red_reference_sum
  • dark_reference_sum
  • green_reference_sum
  • flags
  • blue_reference_sum
  • red_sample_count
  • mass_flow
  • green_sample_count
  • dark_signal_sum
  • mass_flow_output
  • blue_sample_count
  • dark_sample_count
  • red_signal_sum
  • green_signal_sum
  • blue_overflow_count


Back To Table of Contents

DQRID : D161211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/15/2016132911/17/20161929Incorrect
Subject:
OLI/PSAP/M1 - Incorrect reset of transmittance after filter change
DataStreams:oliaospsaprM1.a0
Description:
Operator MAY have changed filter but didn't reset transmission to 100%.  Data invalid.
Measurements:oliaospsaprM1.a0:
  • red_overflow_count
  • blue_signal_sum
  • green_overflow_count
  • dark_overflow_count
  • red_reference_sum
  • dark_reference_sum
  • green_reference_sum
  • flags
  • blue_reference_sum
  • red_sample_count
  • mass_flow
  • green_sample_count
  • dark_signal_sum
  • mass_flow_output
  • blue_sample_count
  • dark_sample_count
  • red_signal_sum
  • green_signal_sum
  • blue_overflow_count


Back To Table of Contents

DQRID : D161211.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/26/2016090011/26/20161022Incorrect
11/11/2016112411/11/20161639Incorrect
11/26/2016110011/26/20161208Incorrect
11/19/2016120911/19/20161637Incorrect
11/30/2016110111/30/20161622Incorrect
11/27/2016181311/28/20161633Incorrect
11/15/2016133011/17/20161929Incorrect
11/01/2016090811/03/20161622Incorrect
11/12/2016121611/12/20161652Incorrect
Subject:
OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, power failures
DataStreams:oliaospsaprM1.a0
Description:
The manufacturer specifies filter must be changed for transmission less than 0.7.  This is 
not always possible with Operator staffing.  

This is different from D161211.1 where on that DQR, the filter was changed, but 
transmittance not reset.
Measurements:oliaospsaprM1.a0:
  • red_overflow_count
  • blue_signal_sum
  • green_overflow_count
  • dark_overflow_count
  • red_reference_sum
  • dark_reference_sum
  • green_reference_sum
  • flags
  • blue_reference_sum
  • red_sample_count
  • mass_flow
  • green_sample_count
  • dark_signal_sum
  • mass_flow_output
  • blue_sample_count
  • dark_sample_count
  • red_signal_sum
  • green_signal_sum
  • blue_overflow_count


Back To Table of Contents

DQRID : D161211.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/13/2016203010/15/20161220Incorrect
10/02/2016114110/02/20161544Incorrect
10/06/2016114410/07/20160749Incorrect
10/30/2016073010/30/20161715Incorrect
10/27/2016231110/28/20161617Incorrect
10/10/2016092610/10/20161551Incorrect
10/23/2016211410/23/20162238Incorrect
10/26/2016075110/26/20161609Incorrect
Subject:
OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off, etc.
DataStreams:oliaospsaprM1.a0
Description:
Manufacturer specifies transmission greater than 0.7 for valid data.  This isn't always 
possible with Operator staffing.
Measurements:oliaospsaprM1.a0:
  • red_overflow_count
  • blue_signal_sum
  • green_overflow_count
  • dark_overflow_count
  • red_reference_sum
  • dark_reference_sum
  • green_reference_sum
  • flags
  • blue_reference_sum
  • red_sample_count
  • mass_flow
  • green_sample_count
  • dark_signal_sum
  • mass_flow_output
  • blue_sample_count
  • dark_sample_count
  • red_signal_sum
  • green_signal_sum
  • blue_overflow_count


Back To Table of Contents

DQRID : D161211.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/20/2016102209/20/20161615Incorrect
09/16/2016180209/16/20162012Incorrect
09/12/2016224509/13/20160016Incorrect
09/22/2016034309/22/20161532Incorrect
09/01/2016000009/01/20161705Incorrect
09/21/2016131309/21/20161612Incorrect
09/16/2016042009/16/20161611Incorrect
09/29/2016062109/29/20161644Incorrect
09/19/2016233409/20/20160058Incorrect
09/28/2016215109/29/20160420Incorrect
09/20/2016074309/20/20161022Incorrect
09/29/2016042009/29/20160621Incorrect
09/25/2016184609/26/20161614Incorrect
Subject:
OLI/PSAP/M1 - Data Gaps > 1h due to Tr<0.7, pump turned off
DataStreams:oliaospsaprM1.a0
Description:
Vendor specifies transmission greater than 0.7 for valid data.  This isn't possible with 
Operating staffing.
Measurements:oliaospsaprM1.a0:
  • red_overflow_count
  • blue_signal_sum
  • green_overflow_count
  • dark_overflow_count
  • red_reference_sum
  • dark_reference_sum
  • green_reference_sum
  • flags
  • blue_reference_sum
  • red_sample_count
  • mass_flow
  • green_sample_count
  • dark_signal_sum
  • mass_flow_output
  • blue_sample_count
  • dark_sample_count
  • red_signal_sum
  • green_signal_sum
  • blue_overflow_count


Back To Table of Contents



END OF DATA