Data Quality Reports for Session: 182600 User: djhackel Completed: 01/04/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D150522.2SGP/AERI/C1 - Cryocooler failuresgpaeriC1.00Suspect
D150729.3SGP/AERI/C1 - Instrument run in slow sample mode.sgpaeriC1.00Does not affect quality
D151010.1SGP/AERI/C1 - Reprocess: Spectral calibration to changesgpaeriC1.00Incorrect
D160318.6SGP/AERI/C1 - Reprocess: intermittent interferogram spikes.sgpaeriC1.00Suspect
D160524.5SGP/AERI/C1 - Radiance contaminated from optical trailer airsgpaeriC1.00Does not affect quality


DQRID : D150522.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/17/2015000004/29/20152246Suspect
Subject:
SGP/AERI/C1 - Cryocooler failure
DataStreams:sgpaeriC1.00
Description:
During this time period the cryocooler began to fail and was eventually replaced.
Measurements:sgpaeriC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D150729.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/18/2015000007/20/20152359Does not affect quality
Subject:
SGP/AERI/C1 - Instrument run in slow sample mode.
DataStreams:sgpaeriC1.00
Description:
Data is fine but the instrument was run in slow sample mode for these 3 days.  We had 
changed the digital meter followed by a short 3rd body test that is run in slow sample mode.  
The instrument wasn't not switched back to rapid sample mode during this period.
Measurements:sgpaeriC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D151010.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/22/2015000008/04/20152359Incorrect
Subject:
SGP/AERI/C1 - Reprocess: Spectral calibration to change
DataStreams:sgpaeriC1.00
Description:
The AERI interferometer laser was changed in the field causing the spectral calibration to 
change. Effects will be more noticeable for higher wavenumber data. Unless doing trace 
gas retrievals likely won't notice much of a change once the reprocessed data is in place.
Measurements:sgpaeriC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160318.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/23/2015000002/19/20162359Suspect
Subject:
SGP/AERI/C1 - Reprocess: intermittent interferogram spikes.
DataStreams:sgpaeriC1.00
Description:
The AERI spectra exhibit a ringing spectral feature on an intermittent basis. The symptom 
is a 'sawtooth' pattern in the spectrum.  This occurs predominantly in Ch2, but also in 
Ch1.  Affected records are unusable and mentor team is going to identify the spikes, 
remove them from the raw data and resubmit raw data to the archive.
Measurements:sgpaeriC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D160524.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/07/2011000002/24/20162359Does not affect quality
Subject:
SGP/AERI/C1 - Radiance contaminated from optical trailer air
DataStreams:sgpaeriC1.00
Description:
Under certain conditions, the observed atmospheric radiance at the lowest level (near the 
surface) is contaminated by air from inside the optical trailer. This is a result of an 
inadequate seal between the AERI front-end (outside) and back-end (inside the optical 
trailer), which let inside air penetrate into the first ~1 m of the AERI optical path.  This 
primarily affects the 540-700 cm-1 region, and retrievals and temperature and water 
vapor.  The result is an artifact in the amplitude of the atmospheric radiance that closely 
mirrors the temperature fluctuations inside the optical trailer, specifically oscillations 
with a period of 10-15 mins.
Measurements:sgpaeriC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents